26
ON/OFF/
ZERO
+
UNIT
Normal weighing mode
(more than 3s)
Enter setup mode when sealed
calibration switch is on?
ON/OFF/
ZERO
+
TARE
Normal weighing mode
(more than 3s)
Enter calibration mode when sealed
calibration switch is on?
Ⅺ
.
Parameters setting summary:
Para-
meter
x/xy
Factory
Set
Setting
Final
Set
P1.xy
00-15
05
Auto-off time: no auto-off; 01-15
minutes auto-off time
P2.x
0, 1, 2
2
0=only Hold function; 1=only Print
function; 2=Print and Hold function,
P3.xy
0,1,2,
3-50
2
0=no hold function; 1=hold lager weight
reading; 2=auto release hold function
when weight is below 10d and auto-hold
new stable weight (more than 10d);
3-50=unchangeable reading when the
variety is 3~50d
P4.x
0,1,2,3,4,
5,6,7
2
0=no RS232 function; 1=output display
data when PRINT pressed; 2=output
gross, tare and net weight when PRINT
pressed; 3=continuously output display
data; 4= continuously output gross, tare
and net weight; 5=output display data
one time when scale is stable; 6=output
gross, tare and net weight one time when
scale become stable; 7=Bio-RS232
compatible (8 data) to NCI-SP1;
P5.x
0, 1, 2, 3,
4
3
Baud rate for RS232: 0=1200bps,
1=2400bps, 2=4800bps, 3=9600bps,
4=19200bps
27
P6.x
0, 1, 2
0
RS232 format: 0=8N1, 1=7O1, 2=7E1
P7.xy
00-31
13
Resolution select: 500,600,750,800,
1000,1200, 1500,2000,2400,2500, 3000,
3500, 4000, 5000, 6000, 7000, 7500,
8000, 10000, 12000, 15000, 20000,
25000, 30000, 35000, 40000, 50000,
60000,70000,75000,80000,100000
P8.x
0,1, 2
0
Division select: 0=1, 1=2, 2=5
P9.x
0,1,2,3,4,
5
0
Decimal point in calibration: 0= x1, 1=
x0.1, 2= x0.01; 3= x0.001; 4= x0.0001;
5= x10
P10.x
0, 1
1
Calibration unit: 0=kg, 1=lb
P11.x
0, 1, 2, 3,
4, 5, 6
6
Weighing units enable: 0=only kg;
1=only lb; 2=only lb:oz; 3=kg or lb;
4=kg or lb:oz; 5=lb or lb:oz; 6=kg, lb, or
lb:oz
P12.x
0, 1, 2, 3,
4, 5, 6, 7,
3
Power-on zero-point range:
0=calibration zero -point +1%FS;
1=calibration zero -point +2%FS;
2=calibration zero-point +5%FS;
3=calibration zero-point +10%FS;
4=calibration zero-point +20%FS;
5=calibration zero-point +50%FS;
6=calibration zero-point +100%FS;
7=No limitation
P13.x 0,1….14
4
Zero range for ZERO button:
0= Power-on zero-point
±
1%FS;
1= Power-on zero-point
±
2FS;
2= Power-on zero-point
±
3%FS;
3= Power-on zero-point
±4
%FS;
4= Power-on zero-point
±5
%FS;
5= Power-on zero-point
±1
0%FS;
6= Power-on zero-point
±2
0%FS;
7= No limitation
8= Power-on zero-point +1%FS
9= Power-on zero-point +2%FS
10=Power-on zero-point +3%FS