Below is the drawing of position of the zero-point potentiometer on PCB for 
conference, decrease ADC data by rotating clockwise, and increase ADC data 
by rotating counter-clockwise.   

         
 
 
 
 
 
 
 
 
3.

 

Press 

UNIT

 key to select displaying weight inner code or input the inner 

working voltage value. When the “U x.xx” is displayed, the display digit is 
voltage value, and the unit is V. The proper working power voltage is between 
5V and 8V. 

4.

 

Press 

TARE

 key to display filtered or un-filtered weight A/D data;

 

when 

 is 

on, the data is filtered. 

5.

 

Press 

ON/OFF/ZERO

 key to exit this mode and return to normal weighing 

mode.  

 

. WORKING PARAMETER SETUP: 

1.

 

When scale is in normal weighing mode, press and hold down ON/OFF/ZERO   
and UNIT key until “SEtUP” is shown, that means the scale is in SETUP 
mode. 

2.

 

This indictor has 19 kinds of parameters to be selected and setup by this 
function.  

3.

 

During SETUP mode, press 

UNIT

 key to change the flashed digits, and 

HOLD/PRINT

 key to shift the flashed position. Press 

TARE

 key to confirm 

and save the set data and enter next setting. Press 

ON/OFF/ZERO

 key to exit 

this mode.     

4.

 

Display 

1)

 

A.B

: Item A parameters and one digit can be input.   

2)

 

A.BC:

 Item A parameters and two digits can be input 

 

11 

3)

 

P

AB.C

: Item AB parameters and one digit can be input 

4)

 

P

AB.CD

: Item AB parameters and two digits can be input   

5.

 

Detailed setup: 

5.1) P1.xy: Auto-off time xy= 00-15   

xy = 00 : means no auto-off function 
xy = 01-15: means the scale turns off to save power after 01-15 minutes when 

no variety of weight on scale and no key operation. 

 
5.2) P2.x: 

HOLD/PRINT

 key function definition 

x = 0 : only HOLD function 
x = 1 : only PRINT function 
x = 2 : HOLD and PRINT function (short-time click the button once for Print 

function, and hold the button for several seconds for HOLD function) 

 
5.3) P3.xy: Hold function setting 

xy = 0 : No hold function; 
xy = 1 : When the scale becomes stable, it will hold the larger weight reading; 
xy = 2 : When the scale becomes stable, it will automatically display and hold 

the weight reading. When this weight is below 10d, HOLD function 
will automatically inactivated. When the new weight is more than 
10d, the scale will hold the new weight when stable.   

xy=3-50 : When the weight variety is within the setting range +3~50d, the 

weight reading will remain unchangeable.     

 
5.4) P4.x: RS232 mode setting 

x = 0 : No RS232 function. It will not transmit or receive any data although 

the scale is with RS232 hardware. RS232 function can be only 
activated when scale is in normal weighing mode.   

x = 1: Press PRINT/HOLD key, the scale will output the current stable 

displayed weight reading and weight unit, and not receive any data 
from other equipment. The output format is as below: 

 

    <LF>< reading, minus, decimal point, weight unit><CR><EXT> 

x = 2: Press 

PRINT/HOLD

 key, the scale will output the data of stable 

current gross, tare, net weight reading and weight unit. The format is 

Summary of Contents for MS140/300

Page 1: ...MS140 300 Operation Manual Contents subject to change without notice Version 2 1 2012 01 10...

Page 2: ...aux appareils num riques de la Classe A prescrites dans le R glement sur le brouillage radio lectrique edict par le minist re des Communications du Canada Warnings Electrical installation For your pro...

Page 3: ...c device These devices emit RF and can cause unstable scale readings If your scale ever performs poorly try moving the scale to a different room or location Avoid using in condition of heavy vibration...

Page 4: ...tivity 0 3mV V 3mV V must be fit to 0 2uV display grad 2 Input Resistor 80 3 Output Resistor 10 K LOADCELL WIRING PIN 1 RED EXCITATION PIN2 BLACK EXCITATION PIN3 GREEN SIGNAL PIN4 WHITE SIGNAL 3 KEY F...

Page 5: ...ormal weighing mode and hold down time longer than 3s 1 Same meaning with that in 3 1 2 Same meaning with that in 3 1 3 Same meaning with that in 3 1 4 Power off the scale UNIT TARE ZERO HOLD PRINT UN...

Page 6: ...oint dot in calibration unit P9 and capacity s unit in calibration P10 If the setting of FS is more than 199999 regardless of decimal point and weight unit the FS capacity will be shown by first four...

Page 7: ...d data will blink If the indicator gets reasonable data it will calculate and store all parameters in EEPROM And then it will auto reset and display all segments of LCD full capacity like power on aga...

Page 8: ...Detailed setup 5 1 P1 xy Auto off time xy 00 15 xy 00 means no auto off function xy 01 15 means the scale turns off to save power after 01 15 minutes when no variety of weight on scale and no key ope...

Page 9: ...put compatible 8 data to the NCI SP1 format 5 5 P5 x Baud rate of RS232 communication x 0 1200bps x 1 2400bps x 2 4800bps x 3 9600bps x 4 19200bps 5 6 P6 x RS232 communication data format x 0 8N1 8 di...

Page 10: ...lb Not available 0 0002 lb 0 0001 kg 0 0002 lb Not available 0 002 lb 0 001 kg 0 002 lb Not available 0 02 lb 0 01 kg 0 02 lb 0 5 oz 15 0 2 lb 0 1 kg 0 2 lb 5 oz 2 lb 1 kg 2 lb Not available 20 lb 10...

Page 11: ...power off zero point as power on zero point and power off tare weight as current tare weight 5 16 P16 x Zero tracking range x 0 0d means no tracking x 1 0 25d x 2 0 5d x 3 1d x 4 1 5d x 5 2d x 6 3d x...

Page 12: ...f the zero point is error it will return nine _ characters b The character will be for negative weight or a space character for positive 19 weight Minus sign follows after the first digit c Useless le...

Page 13: ...ip of capacity and the setting of P7 P8 and P9 Table6 Capacity unit is kg or lb Resolution set by P7 Division set by P8 and P9 P8 0 0 0001 0 001 0 01 0 1 1 10 500 0 0500 0 500 5 00 50 0 500 5000 600 0...

Page 14: ...160000 1600000 100000 20 0000 200 000 2000 00 20000 0 200000 2000000 23 Table8 Capacity unit is kg or lb Resolution set by P7 Division set by P8 and P9 P8 2 0 0005 0 005 0 05 0 5 5 50 500 0 2500 2 50...

Page 15: ...as following drawings 1 Placed vertically 2 Placed horizontally 25 Key Definition summary KEY MODE DEFINITION HOLD PRINT Normal weighing mode Enter or exit HOLD mode output the data as per P4 P5 P6 s...

Page 16: ...x 0 1 2 3 4 3 Baud rate for RS232 0 1200bps 1 2400bps 2 4800bps 3 9600bps 4 19200bps 27 P6 x 0 1 2 0 RS232 format 0 8N1 1 7O1 2 7E1 P7 xy 00 31 13 Resolution select 500 600 750 800 1000 1200 1500 200...

Page 17: ...play 0 P16 x 0 1 2 3 4 5 6 7 8 8 Zero tracking range 0 0d no tracking 1 0 25d 2 0 5d 3 1d 4 1 5d 5 2d 6 3d 7 4d 8 5d P17 x 0 1 2 3 2 Data filter intensity 0 very weak 1 weak 2 middle 3 strong P18 x 0...

Page 18: ...845 246 6717 Fax 44 0 845 246 6718 Email sales brecknellscales co uk Web site www brecknellscales co uk USA Brecknell 1000 Armstrong Drive Fairmont MN 56031 Toll Free 800 637 0529 Tel 507 238 8702 Fa...

Reviews: