88 How to Use the Vivo 45
Vivo 45 User Manual
Doc.006149 En Y-2
Installing the FiO
2
Sensor
Calibrating the FiO
2
sensor
The FiO
2
sensor should be calibrated when first used and then at least
once a month.
1
Place the Vivo 45 so the bot-
tom is accessible.
2
Remove the hatch for FiO
2
sensor.
3
Insert the FiO
2
sensor with the
electrical contact side in.
4
Reinstall the hatch.
5
Calibrate the FiO
2
sensor.
When installed, the Vivo 45 automatically detects the sensor, also
after powering off/on and after power failure.
FiO
2
calibration can be performed from the “FiO
2
/CO
2
Calibration” page under the
“Others” section
.