![Breas Medical Vivo 45 User Manual Download Page 152](http://html1.mh-extra.com/html/breas-medical/vivo-45/vivo-45_user-manual_2806367152.webp)
152 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
Internal/Click-In Battery Hot Alarm
I
TEM
D
ESCRIPTION
Definition
An alarm for Internal/Click-In Battery Over-
heat in Discharge will be given when the
internal or click-in battery reaches 55°C
(131°F).
The battery electronics by manufacture stops
discharge at 60°C (140°F).
Priority
High
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication