182 Technical Specifications
Vivo 45 User Manual
Doc.006149 En Y-2
* BTPS (Body Temperature and Pressure Saturated).
% in TgV
0 to 100%.
±1%
Total Rate
0 to 60 bpm.
±1 bpm
Spont Rate
0 to 60 bpm.
±1 bpm
% Spont
0 to 100%.
Not applicable
SpO
2
70 to 100%.
±3 digits.
No motion and flex
sensor.
Pulse Rate
18 to 300 bpm.
±3 digits.
No motion and flex
sensor.
I:E
1:10 to 10:1.
±0.1 unit for E < 9.9,
±1 unit otherwise.
Insp. Time
0.3 to 5 s.
±0.1 s
Rise Time
0.1 to 5 s.
±10% or ±0.1 s, which-
ever is greatest
EtCO
2
0 to 25%.
0 to 15%: ±(0.3 vol% +
4% of reading).
15 to 25%: unspecified
InspCO
2
0 to 25%.
0 to 15%: ±(0.3 vol% +
4% of reading).
15 to 25%: unspecified
P
OWER
S
UPPLY
S
PECIFICATION
AC supply
100 to 240 V AC, tolerance: +10%/-20%,
50 to 60 Hz, 1.0 - 2.0 A.
External DC
19 V DC, tolerance: 19 V ± 6 V.
Max 90 W.
Click-in battery
Capacity: 65Wh. LiIon. Operational time
6.5 hours, lifetime 3 years.
M
ONITORED
V
ALUE
R
ANGE
A
CCURACY
Summary of Contents for Vivo 45
Page 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Page 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Page 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Page 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Page 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Page 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Page 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Page 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...