Trouble-shooting

 

Error message

Situation

Solution

 

No lens filter is attached.

Attach new, clean lens filter.

The thermometer cannot identify 
a secure position of the probe. 
An accurate measurement was 
not possible.

POS = position error

IRT 4020: Clear the display by 
pushing the «start» button once. 
IRT 4520: Clear the display by 
pushing the «I/O» button once.  
Take care that the positioning of 
the probe is correct and remains 
stable.

Ambient temperature is not within 
the allowed operating range 
(10 – 40 °C or 50 – 104 °F).

Allow the thermometer to remain 
in a room for 30 minutes where 
the temperature is between 
10 and 40 °C or 50 and 104 °F.

Temperature taken is not within 
typical human temperature range 
(34 – 42.2 °C or 93.2–108 °F).

HI = too high

LO = too low

Make sure a new, clean lens filter 
is attached and thermometer is 
properly inserted. Then, take a 
new temperature.

System error – self-check display 
flashes continuously and will not 
be followed by the ready beep 
and the ready symbol.

If error persists,

If error still persists,

Wait 1 minute until the thermo-
meter turns off automatically, then 
turn on again.

... reset the thermometer by re-
moving the batteries and putting 
them back in.

… have the thermometer checked 
at a Kaz Customer Service Center.

 

6022353_IRT_AP_N  Seite 10  Donnerstag, 20. Dezember 2007  2:34 14

Summary of Contents for ThermoScan IRT 4020

Page 1: ...IRT 4520 IRT 4020 Type 6022 Type 6023 IRT 4520 4020 EK KURTZ DESIGN 12 02 03 ThermoScan 1 800 327 7226 Type 6022 I O mem 6022351_IRT_CE_S1 Seite 1 Mittwoch 10 Oktober 2007 2 52 14 ...

Page 2: ...20 IRT 4020 9 11 1 800 327 7226 Type 6022 E x a c T e m p Therm oS can ThermoScan xacTemp I O mem 5 4 2 1 3 1 800 327 7226 Type 6022 start 7 8 9 ThermoScan xacTemp start 6022353_IRT_AP_S3 Seite 1 Donnerstag 20 Dezember 2007 2 50 14 ...

Page 3: ...rk Braun ThermoScan measures the infrared heat generated by the eardrum and surrounding tissues To avoid gross temperature dif ferences the sensor itself is brought to a temperature close to that of the human body When the Braun ThermoScan is placed in the ear it continuously monitors the infrared radiation The measurement will be finished and the result is displayed when an accurate reading can b...

Page 4: ...ThermoScan Practice taking temperatures on yourself and healthy family members to determine their normal temperature ranges Note When consulting your physician communicate that the ThermoScan temperature is a temperature measured in the ear and if possible note the individual s normal ThermoScan temperature range as additional reference Normal ranges by site Axillary 34 7 37 3 C 94 5 99 1 F Oral 3...

Page 5: ...clean lens filter 1 is in place before each measurement 2 IRT 4020 Push the start button 8 IRT 4520 Push the I O button 6 During an internal self check the display shows all segments Then the last temperature taken will be displayed together with MEM Wait for the ready signal beep and the ready symbol in the display 3 Fit the probe snuggly into the ear canal then push the start button 8 start IRT ...

Page 6: ...20 Clear the display by pushing the start button once IRT 4520 Clear the display by pushing the I O button once Wait for the ready signal Fit the probe snuggly into the ear canal then push the start button The Braun ThermoScan ear thermometer turns off automatically after 60 seconds of inactivity The IRT 4520 can also be turned off by pressing the I O button for at least three seconds The display ...

Page 7: ...ometer will be switched off Each further pressing of the I O button displays the remaining memory numbers up to MEM 8 MEM 1 is the latest reading MEM 8 would be the oldest The memory mode is left automatically after displaying the oldest reading or after pressing the I O button for at least 1 second Changing the temperature scale Your Braun ThermoScan is shipped with the Celsius C temperature scal...

Page 8: ...pletely dried out you can put a new lens filter on and measure If the probe tip is damaged contact Kaz Use a soft dry cloth to clean the thermometer display and exterior Do not use abrasive cleaners Never submerge this thermometer in water or any other liquid Store thermometer and lens filters in a dry location free from dust and contamination and away from direct sunlight Additional lens filters ...

Page 9: ... a room for 30 minutes where the temperature is between 10 and 40 C or 50 and 104 F Temperature taken is not within typical human temperature range 34 42 2 C or 93 2 108 F HI too high LO too low Make sure a new clean lens filter is attached and thermometer is properly inserted Then take a new temperature System error self check display flashes continuously and will not be followed by the ready bee...

Page 10: ... until it snaps in place To protect the environment dispose of empty batteries at your retail store or at appropriate collection sites according to national or local regulations Calibration The thermometer is initially calibrated at the time of manufacture If this thermometer is used according to the use instructions periodic re adjustment is not required If at any time you question the accuracy o...

Page 11: ...L EQUIPMENT Please do not dispose of the product in the household waste at the end of its useful life Disposal can take place at your local retailer or at appropriate collection points provided in your country Guarantee Kaz grants 2 years guarantee on the product commencing on the date of purchase Within the guarantee period Kaz will eliminate free of charge any defects in the appliance resulting ...

Reviews: