![Braun SpaceCom Service Manual Download Page 105](http://html1.mh-extra.com/html/braun/spacecom/spacecom_service-manual_2804983105.webp)
Procedural Instructions on the TSC
8
SpaceStation, 2.0 gb
8 - 3
2.0
Housing Leakage Current
(with the SpaceStation with SpaceCom)
Measurement is to be carried out at nominal v10% in sin-
gle fault condition (SFC = Single Fault Condition) incl. mains ca-
ble.
1.
Measure the housing leakage current between mains input at
the mains cable and the metal frame of the Ethernet (RJ 45)
connection in the SpaceCom plug strip.
Note
The metal frames of the connectors in the SpaceCom plug strip
are connected with each other.
2.
Document value.
Abb.: 8 - 1
Legende zu Abb. 8 - 1:
ItemDesignation
1
Metal frame of the Ethernet (RJ 45) connection
Summary of Contents for SpaceCom
Page 1: ...Service Manual SpaceStation SpaceCover SpaceCom Revision 2 0 English 0 2 0 ...
Page 12: ...Contact Persons 0 0 12 SpaceStation 2 0 gb For your information 2 0 ...
Page 34: ...System Overview 1 1 22 SpaceStation 1 0 gb 1 0 For your information ...
Page 42: ...Unit Diagnosis Calibration 2 2 8 SpaceStation 2 0 gb 2 0 For your information ...
Page 68: ...Disassembly Assembly SpaceStation 3 3 26 SpaceStation 1 0 gb 1 0 For your information ...
Page 84: ...Disassembly Assembly SpaceCom 4 4 16 SpaceStation 2 0 gb 2 0 For your information ...
Page 98: ...Disassembly Assembly SpaceCover 5 5 14 SpaceStation 1 0 gb 1 0 For your information ...
Page 100: ...Servicing the Unit 6 6 2 SpaceStation 1 0 gb For your information 1 0 ...
Page 108: ...Procedural Instructions on the TSC 8 8 6 SpaceStation 1 0 gb 1 0 For your information ...
Page 110: ...Test Equipment and Special Tools 9 9 2 SpaceStation 1 0 gb For your information 1 0 ...
Page 118: ...Spare Parts List 10 10 8 SpaceStation 1 0 gb 1 0 ...
Page 120: ...Revision Documentation 11 11 2 SpaceStation 1 0 gb 1 0 ...
Page 124: ...11 11 6 SpaceStation 1 0 gb 1 0 ...