Procedural Instructions on the TSC
8
8 - 2
SpaceStation, 2.0 gb
Earth Leakage Current
Note
Due to the technical design the patient- and housing leakage cur-
rent is not to be measured on the SpaceStation without Space-
Com and without pump inserted. The earth leakage current is
always measured as it is the upper limit of leakage currents which
can be measured.
The earth leakage current of each SpaceStation must be measured
separately.
Check
1.
Before carrying out the measurement remove all interface
cables to the other units.
2.
Remove all pumps out of the SpaceStation.
3.
Measure the earth leakage current without single fault con-
dition (NC = Normal Condition) incl. mains cable.
4.
Document value.
5.
Measure the earth leakage current in single fault condition
(SFC = Single Fault Condition) incl. mains cable.
6.
Document value.
2.0
Summary of Contents for SpaceCom
Page 1: ...Service Manual SpaceStation SpaceCover SpaceCom Revision 2 0 English 0 2 0 ...
Page 12: ...Contact Persons 0 0 12 SpaceStation 2 0 gb For your information 2 0 ...
Page 34: ...System Overview 1 1 22 SpaceStation 1 0 gb 1 0 For your information ...
Page 42: ...Unit Diagnosis Calibration 2 2 8 SpaceStation 2 0 gb 2 0 For your information ...
Page 68: ...Disassembly Assembly SpaceStation 3 3 26 SpaceStation 1 0 gb 1 0 For your information ...
Page 84: ...Disassembly Assembly SpaceCom 4 4 16 SpaceStation 2 0 gb 2 0 For your information ...
Page 98: ...Disassembly Assembly SpaceCover 5 5 14 SpaceStation 1 0 gb 1 0 For your information ...
Page 100: ...Servicing the Unit 6 6 2 SpaceStation 1 0 gb For your information 1 0 ...
Page 108: ...Procedural Instructions on the TSC 8 8 6 SpaceStation 1 0 gb 1 0 For your information ...
Page 110: ...Test Equipment and Special Tools 9 9 2 SpaceStation 1 0 gb For your information 1 0 ...
Page 118: ...Spare Parts List 10 10 8 SpaceStation 1 0 gb 1 0 ...
Page 120: ...Revision Documentation 11 11 2 SpaceStation 1 0 gb 1 0 ...
Page 124: ...11 11 6 SpaceStation 1 0 gb 1 0 ...