Pull the aiming sleeve for the stabilization
screw
17, 18
off the centralizer
27a
.
Locking the polyaxiality
►
Snap the guide tube
7
with the counter torque
8
onto the centralizer
27a
in situ.
If you cannot snap the guide tube
7
onto the central-
izer
27a
:
►
To snap the guide tube onto the centralizer:
Remove the screwdriver
6, 13
from the guide
tube
7
.
- or -
►
Guide the guide tube
7
onto the centralizer
27a
in
an orthograde direction.
►
Insert the screwdriver for retaining bolt
13
into the
guide tube
7
.
►
Attach the torque wrench for retaining bolt
21
to
the screwdriver for retaining bolt
13
.
►
Tighten the retaining bolt
27b
.
The retaining bolt is tightened down when two
clicks are heard.
WARNING
Risk of injury if the aiming
sleeve for the stabilization
screw becomes jammed in the
centralizer as a result of severe
angular misalignment between
the approach direction and
centralizer axis!
►
For the controlled removal
of a jammed aiming sleeve
for a stabilization screw,
unscrew it with a 6.5 mm
yellow stabilization screw
at least 40 mm long.
WARNING
Risk of injury and/or damage
to the implant components
through non-observance of
the correct sequence of oper-
ating steps!
►
Only fit the torque wrench
for retaining bolt onto the
screwdriver for retaining
bolt when this is fully
inserted in the retaining
bolt.
►
Do not slide the screw-
driver for retaining bolt in
the guide tube together
with the torque wrench for
retaining bolt, and do not
let it slide down too
quickly.
Summary of Contents for Aesculap MACS II
Page 2: ...1 A 2 ...
Page 3: ...4b 5d 5a 5c 5b 3 4a 4 B 5 ...
Page 4: ...6 7 9 8 C ...
Page 5: ...10 11 7 13 12 8 D ...
Page 6: ...7 11 10 13 12 8 14 E ...
Page 7: ...17 18 19 20 15 6 16 F ...
Page 8: ...13 7 21 8 G ...
Page 9: ...11 13 6 15 9 20 10 22 H ...
Page 10: ...13 7 1 22 8 I ...