11
ACCULAN dermatome GA 643
Removing the rechargeable battery
¾
If the rechargeable battery needs to be changed during operation,
carry out the battery change in sterile conditions only, using the
changing funnel GA 644 for the removal and re-insertion of bat-
teries, see instructions for use GA 644.
¾
The rechargeable battery
7
always has to be taken out after oper-
ation and before processing.
¾
Push down both release buttons
4
on the cover
5
simultaneously.
¾
Remove the cover
5
.
¾
Prior to battery removal, make sure that the rotational speed con-
trol button
2
is in the front position (rest position).
¾
Tap the ACCULAN dermatome
1
on the palm of your hand until the
rechargeable battery
7
comes loose, and remove the battery.
Disassembling the flap rod of the ACCULAN dermatome
For disinfecting and cleaning, the head of the ACCULAN dermatome
has to be disassembled.
¾
Disassemble the flap rod
22
and the flaps
13
:
– Loosen the nut
24
by turning it clockwise (left-handed thread).
– Turn the nut
24
back to the end of the visible surface.
– Push the flap rod
22
sideways by approx. 4 mm and twist it until
it can be removed.
– Remove the flap rod
22
.
¾
Slide off the flaps
13
from the flap rod
22
.
Assembling the flap rod of the ACCULAN dermatome
¾
Before assembling the flaps
13
, lightly spray all components with
Sterilit M.
¾
Position the flaps
13
on the flap rod
22
against the edge of the rear
surface, taking note of the symbol on the face of the thread.
¾
Screw the nut
24
counterclockwise onto the thread of the flap
rod
22
(left-handed thread).
¾
Turn the nut
24
up to the edge of the visible surface.
¾
Insert and twist the complete flap rod
22
in its guide slots
21
on
the side surfaces.
¾
Push the flap rod
22
sideways as far as it goes, so that the cross
pin of the flap rod comes to lie in the guide slot
21
.
¾
Tighten the nut
24
counterclockwise.
4.2 Attaching the accessories
Securing the ACCULAN dermatome
In order to avoid that the ACCULAN dermatome is unintentionally
turned on during a change of blades, the rotational speed control
button
2
can be secured with a safety catch. During normal operation,
the safety catch remains in the ON position.
¾
Rotate the safety catch
3
to the OFF position.
The button is locked and the ACCULAN dermatome cannot be
turned on.
¾
To deactivate the safety catch
3
turn it back to the ON position.
Fig. 2
Inserting a dermatome blade
¾
Push both clamp levers
9
in the direction of the arrows.
The blade cover
10
is now unlocked.
¾
Open the blade cover
10
in the direction of the arrows.
¾
Insert the dermatome blade
8
in the pusher dog
14
and the blade
guide peg
15
.
¾
Close the blade cover
10
.
¾
Tighten both clamp levers
9
until the roll pin
20
snaps into the
notch of the clamp lever
9
.
Removing the dermatome blade
¾
Release both clamp levers
9
.
¾
Open the blade cover
10
.
¾
Remove the dermatome blade
8
from the pusher dog
14
and the
blade guide peg
15
.
4.3 Function check
Always perform a test run before using the ACCULAN dermatome:
¾
Make certain that a rechargeable battery has been installed.
¾
Use only devices that are in perfect condition.
¾
Operate the ACCULAN dermatome for a short while, at a blade
stroke number of approx. 6 300–6 800 strokes/min.
¾
Note the instructions in chapter ”Care and checking for damage”.
Note
Each intraoperative battery change has to be followed by a function
check!
CAUTION
The ACCULAN dermatome will be damaged through
knocking it against hard surfaces!
¾
Only tap the ACCULAN dermatome on the palm
of your hand.
WARNING
Risk of infection or contamination if surgical
gloves are damaged!
¾
Avoid contact with the cutting edge when in-
serting the dermatome blade.
WARNING
Risk of injury and material damage through unin-
tentionally putting the ACCULAN dermatome into
operation while inserting the blade!
¾
Before inserting the blade, activate the safety
catch.