4
Charging
• Ensure handpiece is completely dry before placing in charger.
Charger ready: Steady green light (without handpiece in base)
Charging in progress: Steady red light. Handpiece
will display an alternating dash “-” pattern
Charging complete: Steady green light (while
handpiece in base)
Note: Completely charge battery at least every 2 months to
maintain optimum battery life and performance. If battery needs
to be replaced, contact Brasseler USA.
CLEANING AND STERILIZATION
• Sterilize light guide before fi rst use.
• Sterilize light guide after each patient use.
• Keep in sterilization pouch until ready to use.
• Wear appropriate personal protective equipment during the cleaning and sterilization
process. This should include eye protection and gloves.
• Do not run or submerge handpiece in water.
• Do not sterilize handpiece body.
1. Wipe handpiece surface with a soft cloth soaked in warm water or a mild detergent mixture.
2. Remove light guide, place in sterilization pouch and seal. Bag light guides separately.
3. Dry handpiece completely before returning it to the charger.
TROUBLESHOOTING
PROBLEM
CAUSE
SOLUTION
Display is blank after
pressing the “Start/Stop”
button
Handpiece is not charged
Charge handpiece
Defective battery
Contact Brasseler USA
Integrated circuit board is
defective
Contact Brasseler USA
Handpiece doesn’t charge
DC adapter is not properly
connected
Check outlet and charger base
connections
Handpiece is inserted
backward in charger
Reposition handpiece as shown
in the setup section
Poor connection between
handpiece and charger
Check handpiece and base
charging contacts to make sure
they are clean and debris free
Defective battery
Contact Brasseler USA
Defective charging base
Contact Brasseler USA