S83/85 Ultrasonic Generator
Installation
A40-064 Rev 0
Page 27
3.7.2
Installation Do’s and Don’ts
1. DO make sure you have all necessary system components: the RF cable, the interconnecting
jumper cables, and the junction boxes.
2. DO ensure that the frequency and number of elements (transducers) of the ultrasonic power
supply matches that of the ultrasonic tank load.
3. DO compare the numbers on the RF cable label and the ID plate on the back panel of the
ultrasonic power supply to make sure they match.
4. DO turn off main power switches.
5. DO make sure to ground the ultrasonic power supply in accordance with local building and
safety codes.
6. DO check the line voltage on the ID plate for the ultrasonic power supply line voltage to
ensure it matches your AC source voltage
7. DO use Power Modulation mode – after initially filling the ultrasonic tank – to remove large
obstructive gas bubbles which shield the ultrasonic energy. Run for 10-15 minutes at
operating temperature.
8. DO maintain the liquid level of the ultrasonic tank with 3 inches of the top – the same level as
the fill port.
9. DO, in application with large parts which displace significant amount of liquid, take care not
to expose the heaters.
10. DO make sure to operate an ultrasonic tank from the same ultrasonic power supply.
11. DO ensure parts are at least 2 inches from the radiating surface of the ultrasonic tank or
immersible transducers. Do this by positioning a basket or parts rack to maintain the 2-inch
minimum distance.
12. DO ensure parts slowly agitate up and down while ultrasonic cleaning. This enhances the
cleaning process by flushing away the contaminates as cavitation removes them.
13. DO make sure to open/expose parts to the ultrasonic field as much as possible.
14. DO use baskets constructed of ¼ inch mesh or larger. For small parts, DO use near solid
bottom baskets.
Don’ts
1.
DO NOT place parts directly on the radiating surface of an ultrasonic tank or immersible.
2.
During the cleaning process, DO limit the disturbance of the liquid. Limit flow from fills
or recirculation to 4-5 gallons per minute.
3.
DO NOT ever use highly acidic compounds or flammable chemistries – the C/CH tank
series and immersible transducers are constructed of 316L stainless steel.
4.
DO NOT plug the ultrasonic power supply to AC source until you finish installing the
ultrasonic tank or tank with immersible transducers.
5.
DO NOT operate the ultrasonic power supply with the cover off.
Caution:
Check the model number of the ultrasonic power supply and the model numbers on the
ultrasonic tank that you plan to install.
Equipment damage can occur if the ultrasonic
power supply model number does not match the number of elements in the ultrasonic tank.
Before installing, make sure that you turn off the ultrasonic power supply and disconnect your AC
voltage source. Leave the ultrasonic power supply disconnected until you finish the installation.
Summary of Contents for S83
Page 2: ......
Page 10: ......
Page 12: ......
Page 72: ......
Page 74: ......
Page 82: ...S83 85 Ultrasonic Generator Options A40 064 Rev 0 Page 82 S83 85 Ultrasonic Generator A40 064 ...
Page 95: ...S83 85 Ultrasonic Generator Options A40 064 Rev 0 Page 95 Figure 7 12 OEM I O Installation ...
Page 98: ......
Page 100: ......
Page 104: ...S83 85 Ultrasonic Generator Assistance A40 064 Rev 0 Page 104 ...