background image

 

15

 

9.

 

TROUBLESHOOTING 

This section presents potential AURA™ Monitor operational issues and 

recommended solutions. Contact Branchpoint Technologies using the 

information on the back cover of this manual for additional assistance.

 

TROUBLE CONNECTING WITH THE AURA™ SENSOR 

OR POOR QUALITY 

OF SERVICE 

 

The AURA™ Monitor will display the messages: “Sensor Connection Lost,” 

“Reposition Antenna, Check Connections,” or “Sensor Error” to indicate an 

AURA

TM 

Sensor connection issue. 

 

Reposition the AURA

TM 

Antenna and check connections until a good 

connection is re-

established between the AURA™ Sensor and AURA™

 

Monitor. 

NOTE: 

If the AURA™ Monitor cannot connect to an AURA

TM 

Sensor, try 

repositioning the AURA

TM 

Antenna.

 

TROUBLE CONNECTING WITH THE AURA™ DATA RECEIVER OR POOR 

QUALITY OF SERVICE 

 

The AURA

TM 

Monitor will state “No Receivers Found” if it does not find any

 

receivers ready for connection. In this case, verify that no other AURA

TM 

Monitor is already connected to the target AURA

TM 

Data Receiver. 

 

The AURA™ Monitor will display the words “Pairing Failed,” “Receiver Error,” 

or “Connect Receiver” to indicate failure to connect with a selected 

AURA

TM 

Data Receiver. If the AURA

TM 

Monitor fails to communicate with the AURA

TM 

Data Receiver or determines there is a poor Quality of Service, it will 

automatically disconnect from the AURA

TM 

Data Receiver. 

 

Verify there are no potential sources of interference in close proximity to 

the AURA™ Monitor or 

AURA

TM 

Data Receiver. 

 

Check the connection of the AURA

TM 

Data Receiver to the patient monitor. 

 

After verification, reset the AURA™

 Monitor. 

POINT MEASUREMENT MODE 

In point measurement mode, the patient monitor displays ICP as a series of 

flat line values (10 to 30 seconds at each consecutive averaged value) instead 

of an ICP waveform. The point measurement mode allows the ICP monitoring 

system to operate in a low power state. When in this mode, try re-positioning 

the AURA

TM 

Antenna for better alignment with the AURA

TM 

Sensor. The system 

will automatically transition out of point measurement mode when the issue is 

resolved.

 

ALARMS 

Table 5 lists all possible alarms that can be displayed to the User based on their

 

significance, LED indicator status, and the audio alarm priority.

 

If a medium priority and a low priority alarm are active at the same time, the 

AURA

TM 

Monitor will only sound the medium alarm.

 

If two or more medium alarms are active at the same time, the AURA

TM 

Monitor will cycle through all of them and display the alarm text for each 

alarm as shown in Table 5. 

 

 

Summary of Contents for AURA TK101-A

Page 1: ...TK101 A 2017 Branchpoint Technologies Inc Revised 2018 09 13 Branchpoint Technologies AURATM Monitor Pack INSTRUCTIONS FOR USE...

Page 2: ...This page intentionally left blank...

Page 3: ...RF telemetry from a coupled AURATM Sensor and relays the ICP waveform to a paired AURATM Data Receiver The AURA Data Receiver makes it possible for the data obtained by the AURA Monitor to be display...

Page 4: ...RA Monitor to overheat and may reduce its performance and service life The AURATM Monitor Pack components are only for use in non condensing humidity conditions Do not discard the AURA Monitor in a fi...

Page 5: ...normal or single fault conditions the AURATM Antenna may reach a maximum temperature of 43 C There is a safety cutoff to prevent temperature from exceeding this value and mitigate the likelihood of d...

Page 6: ...urs and the AURA Monitor s functionality is affected attempt to reset the AURA Monitor or contact Branchpoint Technologies for instructions AURA ICP Monitoring System has not been evaluated for pediat...

Page 7: ...ress locking buttons simultaneously to release battery Figure 4 Slide battery to guide markers Figure 5 Lift and remove battery To insert the battery into the AURA Monitor Line up guide markers on the...

Page 8: ...ndicates there is an error with the battery Try to remove then re insert the battery If problem persists replace the battery or contact Customer Service Charge the spare battery when not in use to ens...

Page 9: ...at is not visibly detectable Figure 10 Align the arrow markings on the antenna cable and the AURA Monitor Figure 11 Plug antenna cable into AURA Monitor RUNNING IN PACKAGE CHECK FOR AURA SENSOR Before...

Page 10: ...y No Sensor Found Select the button to re run the in package check and try to move the AURATM Antenna around the marked area again as the AURATM Sensor within the box may slightly shift during transpo...

Page 11: ...URA Antenna securely in place The LED on the AURATM Antenna will illuminate green upon successful connection to an AURATM Sensor Figure 13 Antenna Holster being affixed onto scalp above AURA Sensor CA...

Page 12: ...RECEIVER After pressing to start pairing the AURA Monitor with the AURA Data Receiver the screen will ask the user to plug the AURA Data Receiver into the patient monitor Ensure that the target AURATM...

Page 13: ...The AURATM Monitor will start pairing with the AURATM Data Receiver After successfully paired with the AURATM Data Receiver the AURATM Monitor will display the ICP NOTE If the AURATM Data Receiver is...

Page 14: ...RM LIMITS AND ALARM To enable an alarm when the acquired ICP reaches the high or low limit set by the user select the Menu button to get to the following screen Select Alarm Status and then Enable To...

Page 15: ...m condition detected Audio pattern output repeatedly every 20 seconds until alarm cleared or user mutes the alarm C A C 523 Hz 200ms tone with 4 harmonics A 440 Hz 200ms tone with 4 harmonics space in...

Page 16: ...ee cloth 6 Visually inspect the devices to ensure surfaces are free of residue and soil Inspect for any signs of wear or damage to the devices such as discoloration and cracking If damage is detected...

Page 17: ...RATM Monitor fails to communicate with the AURATM Data Receiver or determines there is a poor Quality of Service it will automatically disconnect from the AURATM Data Receiver Verify there are no pote...

Page 18: ...in a typical medical installation there is no guarantee that interference will not occur in a particular installation If the device does cause harmful interference the user is encouraged to try and c...

Page 19: ...SCLOSURE Branchpoint Technologies has exercised reasonable care in the selection of materials and the manufacture of these products Branchpoint Technologies excludes all warranties whether expressed o...

Page 20: ...g ms Time constants The time constants for full scale deflection of the system with both increasing and decreasing pressure With increasing pressure 10 mmHg 5 75 ms 20 mmHg 5 62 ms 50 mmHg 8 03 ms Wit...

Page 21: ...n condensing Pressure limits Operating Storage 795 mmHg to 700 mmHg 795 mmHg to 375 mmHg RF data transmission AURATM Monitor and AURATM Antenna Frequency range 2 40 2 48 GHz Output power 0dBm 0 001 wa...

Page 22: ...13 567 MHz Software Version BLE SW 0008 rev A MCU SW 0002 rev D AURA BATTERY SPECIFICATIONS Battery Type Li ion rechargeable Capacity 5600 mAh Charge duration Operating 8 hours typical Voltage 11 1 V...

Page 23: ...12 6 VDC per charging bay 895 mA per charging bay Physical dimensions and weight Charger Height Width Depth Weight Charger Power Supply Height Width Depth Weight with cable 5 8 inches 5 8 inches 3 0...

Page 24: ...F Emissions CISPR 11 System Group 2 The AURA Intracranial Monitor System uses Group 2 limits The AURA Intracranial Monitor System must emit electromagnetic energy in order to perform its intended func...

Page 25: ...ypical commercial or hospital environment Surge IEC 61000 4 5 1kV Differential 2kV Common 1kV Differential 2kV Common Mains power quality should be that of a typical commercial or hospital environment...

Page 26: ...Hz 6V1 in ISM be tween 0 15MHz and 80MHz2 80 AM at 1kHz PROFESSIONAL HEALTHCARE FACILITY ENVIRONMENT Radiated RF IEC 61000 4 3 3V m 80MHz 2 7GHZ 80 AM at 1kHz 3V m 80MHz 2 7GHZ 80 AM at 1kHz PROFESSIO...

Page 27: ...820 CDMA 850 LTE Band 5 Pulse modulation2 217 Hz 2 0 3 28 1720 1845 1970 1700 1900 GSM 1800 CDMA 900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation2 217 Hz 2 0 3 28 2450 2400 2750 WLAN 802 11 b...

Page 28: ...not use if package is damaged Use by date Do not reuse single use device Mark for Nationally Recognized testing for safety standards Conforms to AAMI STD ES60601 1 IEC STD 60601 1 6 IEC STD 60601 1 8...

Page 29: ...27 14 ASSOCIATED DEVICE LABELS PN T 0020 AURA Battery Charger Label PN T 0032 AURA Battery Label PN T 0012 AURA Monitor Label PN T 0039 AURATM Antenna Label...

Page 30: ...28 PN T 0019 AURA Monitor Pack Sales Container Label PN T 0022 AURA Monitor Pack Shipping Container Label...

Page 31: ...ction is expressly prohibited by law or create derivative works based upon the Software or Documentation iii copy the Software or Documentation except for back up or archival purposes iv rent lease le...

Page 32: ...BRANCHPOINT MAKES NO GUARANTEES REGARDING THE QUALITY ACCURACY COMPLETENESS EFFECTIVENESS RELIABILITY OR USEFULNESS OF THE SOFTWARE OR RESULTS OBTAINED THEREFROM NO ORAL OR WRITTEN INFORMATION OR ADVI...

Page 33: ...a waiver of or bar the later exercise of that or any other right power privilege or remedy of Branchpoint The terms and conditions contained in this Agreement may not be modified except in a writing d...

Page 34: ...This page intentionally left blank...

Page 35: ...and understanding these instructions CAUTION FEDERAL LAW U S RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN For further information contact Branchpoint Technologies 1 Technology Drive...

Page 36: ...Distributed by Branchpoint Technologies 1 Technology Drive Suite i 811 Irvine CA 92618 United States Phone 1 949 829 1868 MADE IN USA IFU PN T 0027 rev B...

Reviews: