12
Calibrate receiver according to AURA
TM
Data Receiver Instructions For Use.
After calibration for receiver is complete, the AURA
TM
Monitor will
automatically display the ICP.
To unpair an
AURA™
Data Receiver, press the Menu button and choose “Unpair
Receiver” from the
drop-down menu and then confirm unpairing. Unpairing will
also occur when the AURA
TM
Monitor is turned off. Note that if an AURA
TM
Data
Receiver is briefly turned off and on, connection to the prior AURA
TM
Monitor
may be automatically re-established.
SETTING ICP ALARM LIMITS AND ALARM
To enable an alarm when the acquired ICP reaches the high or low limit set by
the user, select the Menu button to get to the following screen. Select “Alarm
Status
” and then “Enable.” To disable the alarm, select “Disable.”
The User can set ICP alarm thresholds. To do so, select ICP High/Low Limit and
use the Up and Down buttons to set the ICP threshold value. The LED light will
flash when the alarm sounds for these thresholds as well.
NOTE:
The factory default ICP low alarm trigger value is 7 mmHg and the
default ICP high alarm trigger value is 20 mmHg. If the User changes these
trigger values, they will persist across power cycles. Trigger values do not
reset to factory settings if the device is turned off and on.