7.4
Cleaning the EM Tracking Unit
Disinfectants
Only use the following disinfectant types to clean the
EM Tracking Unit
:
Disinfectant Type
Example
Alcohol-based
Meliseptol, Mikrozid AF Liquid
Alkylamine-based
Incidin Plus 2%
Active oxygen-based
Perform
Aldehyde/chloride-based
Antiseptica Kombi - Flächendesinfektion
NOTE: Use only surface disinfectants released in your specific market.
NOTE: Surface disinfectants may leave a residue. This can be easily removed using a dry cloth.
How to Clean EM Tracking Unit
Step
1.
Unplug all cables from the
EM Tracking Unit
.
2.
Clean all housing surfaces using a surface disinfectant (see list of Brainlab approved dis-
infectants). Follow disinfectant manufacturer's recommendations.
3.
Carefully clean interfaces, ensuring that no liquids enter the system.
NOTE: Do not spray liquids directly into EM Base Station’s ventilation openings.
CLEANING