STI Robotics OEM 

49703

 

  

Safety Technical Inspection 

 

figure 4: Reference 
points for PE-
resistance test

 

 

figure 5: Reference 
points for leakage 
current test

 

figure 6: Reference 
points for 
insulation 
resistance test 

 

 

 
 

2.

 

Mounting screws (also 
on rear side!) 

3.

 

Blank metal screw-head of side-rail 
mounting screw  

1.

 

Socket(s) of output 
connector 

2.

 

Contacts of mains power 

input connector 

 

M

 

2.

 

PE terminal 

mains 

MD 

2.

 

Blank metal screw-head of side-rail mounting 

screw 

Summary of Contents for 49703

Page 1: ...an Equipment Leakage Current test with Robotics OEM and verify the Equipment Leakage Current from accessible conductive parts complies with the requirements of IEC62353 see figure5 for details The Eq...

Page 2: ...Safety Technical Inspection figure 4 Reference points for PE resistance test figure 5 Reference points for leakage current test v figure 6 Reference points for insulation resistance test 2 Mounting s...

Page 3: ...0 2 liegen 4 Ableitstrom 5 3 3 IEC62353 F hren Sie einen Ableitstromtest mit Robotics OEM durch und stellen Sie sicher dass der Ger teableitstrom von zug nglichen leitf higen Teilen den Anforderungen...

Page 4: ...703 Sicherheitstechnische Pr fung Abbildung 4 Referenzpunkte f r PE Widerstandstest Abbildung 5 Referenzpunkte f r Ableitstromtest v Abbildung 6 Referenzpunkte f r Isolationswiderstandstest 2 Befestig...

Page 5: ...guire un test della corrente di dispersione dell apparecchiatura Robotics OEM e verificare che la corrente di dispersione sia conforme ai requisiti previsti dalla norma IEC62353 per ulteriori informaz...

Page 6: ...st di resistenza di messa a terra di protezione Figura 5 Punti di riferimento per il test della corrente di dispersione v Figura 6 Punti di riferimento per il test di resistenza dell isolamento 2 Viti...

Page 7: ...tics d origine puis v rifiez que le courant de fuite de l quipement par les pi ces conductrices accessibles r pond aux exigences de la norme CEI 62353 voir la figure 5 pour plus de d tails Le courant...

Page 8: ...ts de r f rence pour le test de r sistance du conducteur la terre figure 5 points de r f rence pour le test du courant de fuite v figure 6 points de r f rence pour le test de r sistance d isolement 2...

Page 9: ...compruebe que la corriente de fuga de equipo procedente de los componentes conductores accesibles cumple los requisitos de la norma IEC62353 ver la figura 5 La corriente de fuga del equipo no debe su...

Page 10: ...tos de referencia para la resistencia de la puesta a tierra Figura 5 Puntos de referencia para la prueba de corriente de fuga v Figura 6 Puntos de referencia para la resistencia de aislamiento 3 Torni...

Reviews: