4 Assembly
MN031-543-S0 EN
TissueSeal / TissueSeal PLUS COMFORT
17
4. Assembly
WARNING
Risk of patient injury from non-sterile ligation
instruments!
The ligation instrument is not sterile as delivered. Clean
and sterilize the instrument before using it.
4.1.
Assembling the ligation instrument
1.
With TissueSeal: plug the disposable electrode tip
A1
onto the
holder for jaw
B2
.
5. Operation
5.1.
Before use
The ligation instrument is assembled (see section 4) and prepared (see
section 7, page 20).
WARNING
Risk of patient injury!
Use only approved BOWA ARC generators with ligation
capability (see section 8, page 31).
Use only suitable products and accessories as
described in the system overview.
Use only intact, sterilized products.