NOTE
Risk of handle damage due to compressed air!
Restrict compressed air pressure to 3 bar or less for
handle drying.
4.
Use filtered compressed air to dry the instrument components.
7.5.
Inspection
These products are subject to wear when used as intended,
depending on the intensity of use. This wear arises from the design
and construction of the instruments and is unavoidable.
Replace the product if it shows externally visible defects or does not
function as described in this manual. Please advise the manufacturer
or the manufacturer’s authorized representative in such cases.
After cleaning, visually inspect each of the instruments and test
them for proper operation.
Replace any damaged parts.
We recommend that instrument oil not be used on moving or pivoting
parts, such as the pliers handles. Instrument oil may cause a tendency
to swelling in certain plastics and impair the ease of motion of the
instruments.
DANGER
Risk of patient burns with brittle or defective insulation!
Replace instrument components with damaged
insulation.