11 Technical specifications
900-301_IFU-V2.2_11316-S0-20131023-EN
Operating Manual ARC 250 / 303
61
11.
Technical specifications
Pos : 72 /679-BOWA/HF- Ger äte/ARC 200/250/300e/11 T echnisc he Daten/T ec hnis che M er kmal e @ 7\mod_1308730457223_6.doc @ 52355 @ 2
11.1.
Technical specifications of ARC 250 and ARC
303 HF devices
Insulation type / Classification
EMC
IEC 60601-1-2
Level of protection provided by the housing
IP 21
Protection class according to EN 60601-1
I
Type of application component as specified in
EN 60601-1
CF
Compliance with standards
IEC 60601-1: 2005,
IEC 60601-1-2: 2007,
IEC 60601-2-2: 2009,
ISO 14971: 2007,
ISO 13485: 2003 +
Cor.1_2009
Classification as per EC Directive 93/42/EEC
IIb
Power connection
Power consumption in standby mode
25 VA
Line frequency
50/60 Hz
Maximum power consumption with 300 W HF
output power
500 VA
Connection for potential equalization line
Yes
Voltage range: 230 V
Input voltage range
198 V to 260 V
Current consumption in standby mode
110 mA
Current consumption at maximum HF output
3.0 A
Mains fuses
2 x 3.15 A slow-blow
Voltage range: 100 V, 115 V
Input voltage range
90 V to 130 V
Current consumption in standby mode
220 mA
Current consumption at maximum HF output
6.0 A
Mains fuses
2 x 6.3 A slow-blow
Dimensions and weight
External dimensions: width x height x depth (mm) 430 x 150 x 400
Weight
Approx. 10 kg
Summary of Contents for ARC 250
Page 1: ...OPERATING MANUAL ELECTROSURGICAL UNIT...
Page 2: ......
Page 3: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 3...
Page 4: ...4 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Page 5: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 5...
Page 8: ...Legend 8 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...