6 Operation
44
Operating Manual ARC 250 / 303
900-301_IFU-V2.2_11316-S0-20131023-EN
Pos : 48 /679-BOWA/HF- Ger äte/ARC 200/250/300e/6 Bedi enung/Programm 7 "GastroCUT Pap" (nur mit Opti on GastroC UT) @ 6\mod_1297072563715_6.doc @ 48104 @ 3
6.3.8.
Program 7 "Gastro KNIFE" (only with GastroCut option)
This program is used in gastroenterology. Papillotomes and endoscopic
resection instruments are used for cutting and coagulation.
ARC control technology generates the cutting effect with simultaneously
minimized output power.
ARC control facilitates immediate cutting and prevents adhesion of the
electrodes.
The connection port
29
has no function with the programs
"Gastro LOOP" and "Gastro KNIFE".
Three modes are available in this program:
fast (approximately 2.5 cutting pulses per second),
medium (approximately 1.8 cutting pulses per second) and
slow (approximately 1.4 cutting pulses per second).
For more information on "GastroCut" mode, see section "GastroCut"
mode, page 27.
The key/display "Monopolar Output"
30
is preconfigured.
1. Connect the papillotome to the socket for "Monopolar Output"
30
(use an adapter if necessary).
2. Connect the foot switch to the HF device.
3. Activate the HF current via the yellow key of the foot switch for
cutting.
– or –
Activate the HF current via the blue key of the foot switch for
coagulation.
4. Set the coagulation level using the keys for scabbing level
5/6
.
5. Press keys
3
and
4
to switch between:
slow "Gastro KNIFE" (non-accelerated),
dynamic " Gastro KNIFE" (medium speed; default) and
fast dynamic " Gastro KNIFE" (accelerated).
6. In the case of hemorrhages:
Press the key/display for "Forced Coagulation" current
13
or use a
clip.
7. If necessary, remove the papillotome and connect the argon probe
and the ARC PLUS argon unit.
8. Press the key/display for "Spray Coagulation" current"
14
.
9. Press the blue key of the foot switch for coagulation.
Pos : 49 /679-BOWA/HF- Ger äte/ARC 200/250/300e/6 Bedi enung/M enü- Programme @ 5\mod_1270109501523_6.doc @ 38535 @ 23
Summary of Contents for ARC 250
Page 1: ...OPERATING MANUAL ELECTROSURGICAL UNIT...
Page 2: ......
Page 3: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 3...
Page 4: ...4 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Page 5: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 5...
Page 8: ...Legend 8 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...