3 Functionality
30
Operating Manual ARC 250 / 303
900-301_IFU-V2.2_11316-S0-20131023-EN
3.6.
ARC Control technology
The "ARC Control" regulator takes varying cutting surfaces and cutting
speeds into account, detects differences in tissue and the resulting
changes in resistance, and adjusts the output power to the minimum
level necessary under actual conditions.
Pos : 21 /679-BOWA/HF- Ger äte/ARC 200/250/300e/3 F unkti ons weis e/N eutralel ektr oden-Über wac hung @ 6\mod_1297065725527_6.doc @ 47857 @ 233
3.7.
Neutral electrode monitoring
Always use the largest possible electrode when attaching a neutral
electrode.
3.7.1.
General information
BOWA recommends using split neutral electrodes, since only this type of
electrode allows the HF device to detect detachment of the neutral
electrode if this occurs.
Monitoring of the neutral electrode minimizes the risk of burns at the site
where the neutral electrode is attached.
Two types of neutral electrodes can be monitored:
One-piece neutral electrodes
Split neutral electrodes.
The split neutral electrodes are shown on display
20
, and the one-piece
neutral electrodes are shown on display
19,
see chapter Error Display in
EASY monitoring, page 55.
Summary of Contents for ARC 250
Page 1: ...OPERATING MANUAL ELECTROSURGICAL UNIT...
Page 2: ......
Page 3: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 3...
Page 4: ...4 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...
Page 5: ...900 301_IFU V2 2_11316 S0 20131023 EN Operating Manual ARC 250 303 5...
Page 8: ...Legend 8 Operating Manual ARC 250 303 900 301_IFU V2 2_11316 S0 20131023 EN...