88
LATITUDE
™
Communicator
Patient Manual
Additional Inf
orma
tion
Electromagnetic Emissions and Immunity
Table 2. Guidance and manufacturer’s declaration—
electromagnetic compatibility
The LATITUDE Communicator is suitable for use in a professional health care
facility or home environment.
Test
Compliance
Electromagnetic
environment—guidance
Protection of radio services and
other equipment
CISPR 11
Group 1
Class B
The LATITUDE Communicator
uses RF energy only for its
intended uses in communication
with the implanted device or
connectivity functions. Its RF
emissions are very low and
are not likely to cause any
interference in nearby electronic
equipment.
Protection of the public mains
network
CISPR 11 Class B
IEC 61000-3-2
IEC 61000-3-3
The LATITUDE Communicator is
suitable for use in a professional
health care facility or home
environment.
Electrostatic discharge
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, and
± 15 kV air
Radiated RF EM field
3 V/m from 80 MHz to
2.7 GHz
Proximity fields from RF wireless
communications equipment
380 - 390 MHz: 27 V/m
430 - 470 MHz: 28 V/m
704 - 787 MHz: 9 V/m
800 - 960 MHz: 28 V/m
1700 - 1900 MHz: 28 V/m
2400 - 2570 MHz: 28 V/m
5100 - 5800 MHz: 9 V/m
Rated power frequency magnetic
field
30 A/m
Electrical fast transients/ bursts
± 2 kV input AC power
± 1 kV SIP/SOP
Surges line-to-line
± 0.5 kV, ± 1 kV input AC
power
Surges line-to-ground
± 0.5 kV, ± 1 kV, ± 2 kV input AC
power
Summary of Contents for LATITUDE 6288
Page 1: ...LATITUDE Communicator PATIENT MANUAL ...
Page 2: ......
Page 4: ...LATITUDE Communicator Patient Manual ii ...
Page 32: ...LATITUDE Communicator Patient Manual 24 ...
Page 58: ...LATITUDE Communicator Patient Manual 50 ...
Page 82: ...LATITUDE Communicator Patient Manual 74 ...
Page 101: ......