Message
Description
Error
Code
Possible Cause
Troubleshooting Step
Check the
Connector
Interface Cable
connection to the
Control Unit�
E306
Connector Interface
Cable is disconnected
from the Control Unit
while ultrasound is
running�
1� Ensure the Connector Interface Cable is securely connected to the Control Unit�
2� If the error persists, disconnect and reconnect the CIC�
3� If the error persists, press the Power button to turn off the Control Unit�
4� Let the Control Unit sit for 30 seconds�
5� Use the Power button to turn on the Control Unit�
6� If the error persists, try using another Control Unit 4�0-compatible CIC�
7� If the error persists, please call the phone number listed in the
For
Support
section�
8� If you were able to resume therapy after switching CICs, please call the phone
number listed in the For Support section to troubleshoot the issue with the
original CIC
Replace the
Connector
Interface Cable�
E307
E315
E321
Defective Connector
Interface Cable�
1� Ensure the Connector Interface Cable is securely connected to the Control Unit�
2� If the error persists, disconnect and reconnect the CIC�
3� If the error persists, try using another Control Unit 4�0-compatible CIC�
4� If the error persists, press the Power button to turn off the Control Unit�
5� Let the Control Unit sit for 30 seconds�
6� Use the Power button to turn on the Control Unit�
7� If the error persists, please call the phone number listed in the
For
Support
section�
8� If you were able to resume therapy after switching CICs, please call the phone
number listed in the For Support section to troubleshoot the issue with the
original CIC�
Check the device
connectors�
E311
Poor connection of the
Infusion Catheter and
Ultrasonic Core�
Infusion Catheter
and Ultrasonic Core
Connectors are wet at
the connection site�
1� Disconnect the Infusion Catheter and Ultrasonic Core from the Connector
Interface Cable�
2� Inspect for fluid or damage to the pins�
3� Dry the Infusion Catheter and Ultrasonic Core connectors, if necessary�
4� Reconnect the Infusion Catheter and Ultrasonic Core and ensure the
connections are secure�
5� Press Start to resume ultrasound�
6� If the error persists, disconnect and reconnect CIC�
7� If the error persists, try using another Control Unit 4�0-compatible CIC�
8� If the error persists, please call the phone number listed in the
For
Support
section�
9� If you were able to resume therapy after switching CICs, please call the phone
number listed in the For Support section to troubleshoot the issue with the
original CIC�
Incompatibility
of CIC
E331
The Connector Interface
Cable is not compatible
with the Infusion
Catheter, Ultrasonic
Core, or both�
In rare circumstances
error code E331 may
also be presented due
to a software anomaly�
1� Disconnect the Infusion Catheter and Ultrasonic Core from the Connector
Interface Cable�
2� Inspect for fluid or damage to the pins�
3� Dry the Infusion Catheter and Ultrasonic Core connectors, if necessary�
4� Reconnect the Infusion Catheter and Ultrasonic Core and ensure the
connections are secure�
5� If the error persists, disconnect and reconnect the CIC�
6� If the error persists, try using a Connector Interface Cable that is compatible
with the Infusion Catheter and Ultrasonic Core�
7� If the error persists, please call the phone number listed in the
For Support
section�
The same error
has occurred
three times�
n/a
Control Unit has
detected the same error
three times in a row�
Please call the phone number listed in the
For Support
section