Table 20.
Message
Description
Error
Code
Possible Cause
Troubleshooting Step
Check the
catheter
connector�
E300
E309
E310
E312
E322
Poor connection of the
Infusion Catheter�
Infusion Catheter
Connector is wet at the
connection site�
1� Disconnect the Infusion Catheter�
2� Inspect for fluid or damage to the pins�
3� Dry the Infusion Catheter connector, if necessary�
4� Reconnect the Infusion Catheter and ensure the connection is secure�
5� Resume ultrasound�
6� If the error persists, connect the CIC into the other channel�
7� Resume ultrasound�
8� If the error persists, please call the phone number listed in the For
Support
section�
Please call the
phone number
listed in the
For
Support
section�
E301
E303
Channel error�
1� Disconnect the Infusion Catheter and Ultrasonic Core connectors�
2� Inspect for fluid or damage to the pins�
3� Dry the connectors, if necessary�
4� Reconnect the Infusion Catheter and Ultrasonic Core and ensure the
connection is secure�
5� Press Start to resume ultrasound�
6� If the error persists, disconnect and reconnect the CIC�
7� Press Start to resume ultrasound�
8� If the error persists, please call the phone number listed in the For
Support
section�
Please call the
phone number
listed in the
For
Support
section�
E316
E317
E318
E319
E326
E327
E328
E329
E330
Channel error�
In rare circumstances
error code E319 may
also be presented due
to a software anomaly�
1� Press the Power button to turn off the Control Unit�
2� Let the Control Unit sit for 30 seconds�
3� Use the Power button to turn on the Control Unit�
4� If the error persists, please call the phone number listed in the
For
Support
section�
Please call the
phone number
listed in the
For
Support
section�
E323
E325
Channel error�
1� Disconnect the Infusion Catheter and Ultrasonic Core connectors�
2� Inspect for fluid or damage to the pins�
3� Dry the connectors, if necessary�
4� Reconnect the Infusion Catheter and Ultrasonic Core and ensure the
connections are secure�
5� Resume ultrasound�
6� If the error persists, disconnect and reconnect the CIC�
7� If the error persists, press the Power button to turn off the Control Unit�
8� Let the Control Unit sit for 30 seconds�
9� Use the Power button to turn on the Control Unit�
10� If the error persists, please call the phone number listed in the
For
Support
section�
Please call the
phone number
listed in the
For
Support
section�
E324
Channel error�
1� Make sure nothing is pressing against the Stop and Start buttons�
2� Press Start to resume ultrasound�
3� If the error persists press the Power button to turn off the Control Unit�
4� Let the Control Unit sit for 30 seconds�
5� Use the Power button to turn on the Control Unit�
6� If the error persists, please call the phone number listed in the
For
Support
section�