49
3. Technical data
Measurement chamber
E-chamber principle, exhaust-gas supply at the center
Length
430 mm
Inside diameter
16 mm
Heated; gas temperature in measurement chamber between 70 °C and 200 °C
Measurement
Measurable
Measuring
Resolution Calibration/Road-
variable
range
worthiness limit
Opacity
0 - 100 %
0.1 %
–
Absorption
0 - 10 m
-1
0.01 m
-1
0.3 or 0.45 m
-1
coefficient k in Germany
± 0.15 m
-1
for k
≤
1 m
-1
and
± 0.15 k for k > 1m
-1
in Switzerland
During the calibration procedure, the prescribed measurable variable must be set using
a lead-sealed switch.
Power supply RTM 430
2 x 14.5V~ / 8A
Power supply RTM 430 internal
Processor
+ 12 V
100 mA
- 12 V
100 mA
+ 5 V
200 mA
Ventilator
+ 12 V
1 A
Valve
+ 12 V
1 A
Measurement chamber heater
~ 24 V
4.2 A (100 W)
Dimensions and weight
Dimensions (W x H x D)
594 x 203 x 151 mm
Weight
approx. 8 kg
Noise emission
≤
70 dB (A)
Temperature range
Accuracy
+2°C - +45°C
Accuracy with Winter Kit
-5°C - +45°C
Maximum exhaust-gas temperature
at RTM 430 inlet
200°C