background image

 

8.

 

The status of the A/F channel will be displayed at the bottom of the dialog box.  Before performing a 
calibration, make sure the sensor is not still warming up and that there are no errors.   Do not use the Heater 
Calibration, Restart or Zero Calibration buttons.  If you have installed a new sensor in this channel perform a 
free air calibration.  When you have finished click on the OK button to exit. 

 
 
 

 

 
Free Air Calibration (using a calibrate button): 

It is recommended to periodically perform a free air calibration to correct for sensor wear caused by leaded fuels and 
aging.  Before starting a free air calibration, the sensor must be removed from the exhaust and exposed to open air.  
A free air calibration can be initiated by following the instructions below. 
 

CAUTION:

 Be careful when handling sensors during/after calibration as they will be 

HOT. 

Ensure sensors are 

clear from any items as sensors may burn anything that comes in contact with them. 
 

NOTE:

 If the CALIBRATE button is not readily accessible, you can also perform a free air calibration using the 

DataLink software. 
 

1)

 

Remove each sensor from the exhaust pipes and leave exposed to open air.  

2)

 

Turn on the controller and wait for the status LED to glow solid red. 

3)

 

Press AND

4)

 

After at least 3 seconds,

 release

 the button and within the next 5 seconds, 

press 

 hold

 the CALIBRATE button on the controller for at least 3 seconds. The status LED will 

begin to flash. This confirms the unit is ready to begin the calibration procedure. 

AND

5)

 

Upon successful completion of the calibration process, the status LED will glow solid red. 

 release

 the 

CALIBRATE button once again, to confirm that you want to perform a free air calibration.  The status 
LED will flash rapidly while performing the calibration (only happens for a few seconds).   

6)

 

 Calibration is finished.  

7)

 

Shut off all power, let sensors cool down before handling. 

8)

 

Re-install in exhaust pipes.  

Troubleshooting 

There are two method used to indicate that there is an error on one or more channels. 
 

1.

 

The channel will output a negative value to indicate an error.  If the recorded data shows a 
negative value,  the data is invalid.  

 
2.

 

The LED on top of the controller does not go solid after a few minutes and is flashing quickly, 
about 5 flashes per second.  This indicates that an error has occurred on one or more channels.  

 
 

PROBLEM 

RECOMMENED SOLUTION 

I get a negative reading on one or 
more channels. 

A negative reading indicates an error has occurred.  Look at the error code 
chart below to determine the exact cause of the error. 

The LED on top of the controller 
never stops flashing. 

A continuously fast flashing status LED indicates that an error has occurred 
on one ore more channels.  Perform the calibration procedure outlined 
above on each channel.  The status box will describe the cause of the error. 

I am using the sensors with methanol 
and the reading are very lean. 

The factory default configuration is set to output air fuel ratio units for 
gasoline.  Follow the steps outlined in the ‘Software Setup and Module 
Configuration’ section to set the sensor type to ‘A/F Units, Methanol’ 

 
 
 

IMPORTANT!! 

You must remove the sensor from the exhaust and expose the sensor to open air before performing a free 
air calibration.  After removing the sensor from the exhaust, wave the sensor back and forth a few times to  
be sure all of the exhaust gases are purged from the sensor.  

Summary of Contents for LSU4

Page 1: ...recommended since oxygen can pass back over the sensor during overlap causing lean readings particularly at low engine RPM Weld the sensor bung approximately 12 inches from the header flange Mounting the sensor too close to the exhaust valve can cause the sensor to overheat To avoid sensor damage from condensation do not mount the sensor on the bottom of the exhaust tube Each sensor has been calib...

Page 2: ...ger 2 Turn the power on to the data logger and the air fuel sensor controller 3 Start the DataLink program on your PC 4 Open the data logger configuration file by selecting Open Car Configuration under File from the menu bar The following dialog box will appear 5 Select the configuration that contains your logger type and serial number Example V300_1001 6 Left click on Edit from the menu bar then ...

Page 3: ...nding the Configuration to the module section below Setting the Channel Mode If you are not going to use a sensor with one of the channels for instance you have a four channel controller but are using only three sensors you should set the channel mode for the unused channel to Disabled If you do not disable the channel the LED indicator on the control box will flash an error indicating the sensor ...

Page 4: ...must be resolved before the air fuel sensor s can take data Follow the troubleshooting instructions below Free Air Calibration A free air calibration is also recommend to correct for changes in atmospheric pressure and sensor wear caused by leaded fuels and aging Before starting a free air calibration the sensor must be removed from the exhaust and exposed to open air A free air calibration must b...

Page 5: ... release the CALIBRATE button once again to confirm that you want to perform a free air calibration The status LED will flash rapidly while performing the calibration only happens for a few seconds 6 Calibration is finished 7 Shut off all power let sensors cool down before handling 8 Re install in exhaust pipes Troubleshooting There are two method used to indicate that there is an error on one or ...

Reviews: