41 to 45 IP Module programming steps
Refer to the user manual of the IP Module, these are optional steps.
50 to 58 Alarm type links
It is possible to link an alarm type to specific call numbers. Each
programming step refers to a certain alarm type.
–
50
: wireless transmitter (emergency call with a wireless transmitter)
–
51
: emergency call (with the Carephone 62)
–
52
: fire / intrusion
–
53
: repeated emergency call and local confirmation
–
54
: sign in / sign out
–
55
: service call / direct call. The direct call can only made to a
telephone without acknowledgement. A single attempt will be made.
–
56
: technical messages 1 (power failure, power restored, unit battery
failure, unit battery low, line failure, line restored)
–
57
: technical messages 2 (automatic test call, radio jamming, radio
transmission monitoring, transmitter battery low)
Choose which destination numbers
1
to
10
are associated with each alarm
type (call sequence). If no destination number is entered, then all
programmed call numbers will be called. It is not possible to have the
same destination number twice.
–
58
: registration call / presence marking - service done
Registration call: after an emergency call, the destination number
entered in setting 58 will be called for registration purposes. The call
destination must be a monitoring centre.
Presence marking - service done: see step
27
.
Choose which call numbers
1
to
10
are associated with this alarm type
(call sequence). If no destination number is entered, then no call is made.
It is not possible to have the same destination number twice.
60 Number of announcements when calling a telephone
Specify how often the Carephone will announce the recorded message.
–
0
= no announcement, from
1
to
9
for the required number of
announcements. Default value is
2
.