- 9 -
Use the long axis of the scaling tip (insert) to wipe accretions from the tooth. Do not gouge
the tooth with the tip.
3-C: Precautions for
hand-piece
In order for the system’s hand-piece to function consistently and effectively, Accumulated
air pressure from the handle (or hand-piece) must be released every 10 to 15 minutes on
every procedure by doing the following steps:
1. Put the insert into the hand-piece and make sure it is fully seated.
2. Hold the hand-piece vertically with the insert tip on top; depress the foot Switch (control).
3. Let the water spray out for a minute or so, then start scaling procedure.
Section IV: Infection Control
4-A: General Infection Control Recommendations:
As with all dental procedures, use standard protective equipment such as face masks, eyewear, face
shields, gloves and protective gowns.
To ensure safety to the operator and patient, carefully follow the Infection Control Procedures
detailed in this section.
4-B: Water Supply Recommendations
All dental water supply systems should conform to applicable Centers for Disease Control and
Prevention (CDC) and the American Dental Association (ADA) standards. These standards should
apply to flushing, chemical flushing, and general infection control procedures.
4-C: Cleaning and Sterilization:
Hand piece
Prior to cleaning, remove the insert from the hand piece. Next, drain any water left in the
hand piece to clear the hand piece of any residue accumulated during treatment. Then, clean
the surface of the hand piece with antiseptic soap or solution and rinse thoroughly with water.
Afterwards, wipe or spray a chemical disinfectant on the hand piece. Please try to use a
nonabrasive disinfectant to avoid damaging the hand piece.