Summary of Contents for KRUUSE-SP1

Page 1: ...Manual Bonart Co Ltd Jorgen Kruuse A S 4F 11 No 3 Wuquan 1st Rd Xinzhuang Dist New Taipei City 24892 Taiwan R O C Tel 886 2 22983980 Fax 886 2 22983981 Email sales bonartmed com Havretoften 4 DK 5550 Langeskov Denmark 0434 COPYRIGHT 2012 BONART CO LTD ...

Page 2: ...dvertent or deliberate publication BONART intends to enforce its rights to this manual under the copyright law Reproduction of the materials presented in this manual without the express written permission of BONART Co Ltd is prohibited The contents of this manual are subject to change without prior notice PROPERTY OF BONART CO LTD ALL RIGHTS RESERVED ART is a registered trademark of BONART Co Ltd ...

Page 3: ...d and properly trained personnel NOTE CAUTION AND WARNING STATEMENTS NOTE Provides tips and advice CAUTION Provides correct operating or maintenance procedures WARNING Alerts users of danger that may cause severe injury when proper procedures are not followed SYMBOLS Complies with MDD 93 42 EEC BF equipment Grounding terminal Attention Please read instructions ...

Page 4: ...emakers Patients with pacemakers should avoid treatment with the KRUUSE SP1 Electronic appliances including razors hair dryers microwave ovens TV receptors and other medical equipment such as the KRUUSE SP1 may interfere with the performance of pacemakers For more information on this subject please refer to the following articles Advances in Cardiac Pacemaker The New York Academy of Sciences Vol 1...

Page 5: ...lted or unstable surface may degrade performance or damage the device Do not modify the device Modifications to the device will violate safety codes and invalidate the warranty while endangering the patient and operator Do not place heavy objects on top of the power cord and keep the power cord away from intense heat Unplug the device and call Bonart for service and instructions if you observe any...

Page 6: ...lexes and moves an installed scaling tip insert approximately 25 000 cycles per second in an elliptical pattern based on the amount of power used Thus more power results in greater tip displacement which allows for more robust scaling However the vibrating metal stack generates heat and requires adequate amounts of fluid to cool the hand piece and tissues during treatment When the two parts of the...

Page 7: ...f stainless steel avoid direct contact with teeth Direct contact will cause damage to teeth enamel For optimum performance we suggest practicing on a model or aluminum plates to become more familiar with the scaling tips before performing on patients ...

Page 8: ...on II Contraindications and Warnings 8 Section III Precautions 8 Section IV Infection Control 9 Section V Installation 11 Section VI General Description and Information of Parts 15 Section VII Techniques 18 Section VIII System Maintenance 19 Section IX Troubleshooting 20 Section X Specifications 21 Section XI Disposal 22 ...

Page 9: ...ts and Systems Ensure sufficient water flow to the scaler tip during use to cool the hand piece and insert Take precaution as all ultrasonic scalers produce aerosols that may transmit contagious diseases Keep the unit away from intense heat Excessive heat may damage the electronic components Switch the water valve off when the KRUUSE SP1 is not in use Avoid treating patients with pacemakers due to...

Page 10: ...nt such as face masks eyewear face shields gloves and protective gowns To ensure safety to the operator and patient carefully follow the Infection Control Procedures detailed in this section 4 B Water Supply Recommendations All dental water supply systems should conform to applicable Centers for Disease Control and Prevention CDC and the American Dental Association ADA standards These standards sh...

Page 11: ...rt with a brush water and detergent Be sure to rinse thoroughly after scrubbing Let the insert dry Then place the insert in a bag and autoclave at 275 135 for 15 minutes or as recommended by the manufacturer of your sterilization unit to sterilize the insert Bio indicators or chemical indicators for ensuring the efficiency of the sterilization cycle are recommended NOTE Routine spore testing will ...

Page 12: ...ngle should be cleaned to remove hair and debris after each patient At the end of the day it should be cleaned and lubricated To clean the Prophy Head follow the steps below 1 Remove the rubber cup 2 Remove the cup holder with pliers by turning the knurled knob right the knurled knob has a reversed thread Wash away debris from the inside Then put 1 drop of oil onto the gears 3 in 1 multipurpose oi...

Page 13: ... 12 Please use the manual shut off valve on the dental water system supply line to shut off water when the office is unoccupied 5 C Electrical Requirements Refer to Section X Specifications ...

Page 14: ...F 50 Streamline Universal Plastic Handle Internal Flow Insert ART P 10 Straight Angle Universal Metal Handle External Flow Insert Optional Optional 7 Water Hose Quick Disconnect Assembly Optional 8 Micro motor 1 9 Micro motor seat 1 10 Straight Hand piece Optional 11 Prophy Angle Optional NOTE While unpacking check your KRUUSE SP1 unit for any damages If any damages are found please contact your d...

Page 15: ...eated 5 G Water Supply Line Connection Push the blue water tube hose into the stainless steel receptacle until the hose cannot be pushed any further Then tighten the screw Connect the quick connect to the water supply line The KRUUSE SP1 should be prepackaged with a male quick connect If a female quick connect is necessary please contact a Bonart distributor to purchase it Inspect all connections ...

Page 16: ...es as a placeholder Remove the rod after you open the lock B Match the groove at the base of the prophy head with the top of the straight hand piece and connect them together C Close the lock by turning the locking ring counter clockwise The prophy head should now be locked onto the straight hand piece and should not fall off during operation D If you wish to remove the prophy head follow step A t...

Page 17: ...d piece 6 Power control knob 10 Power cord socket 3 Power button 7 Micro motor socket 11 Forward Reverse switch 4 Micro motor Mode button 8 Foot switch socket Figure 2 KRUUSE SP1 diagram 1 Main Unit The main unit generates power and produces a signal that is passed to the hand piece The hand piece is then powered up and vibrates the installed insert 2 Hand piece The hand piece is the housing for s...

Page 18: ...or should be lit indicating that the system is operating in polishing mode Under polishing mode the micro motor is engaged To return to scaling mode push the button again 5 Water control knob The water control knob allows you to control the amount of water that flows through the hand piece If turned clockwise the amount of water will decrease If turned counter clockwise the amount of water will in...

Page 19: ...Water hose input connector The water hose input connector is the contact point for the water tube and the main water supply line Use either a male or female quick connect to connect the water tube to the water supply The KRUUSE SP1 comes with a male quick connect attached to the water tube 10 Power cord socket The power cord socket is where you connect the power cord to the main unit 11 Forward Re...

Page 20: ...and piece will melt 5 Lubricate the rubber O ring on the insert with water before placing it into the hand piece Install the insert by firmly pushing until seated 6 Hold the hand piece over a sink or drain with an insert installed Verify that water is reaching the tip 7 Check your scaling inserts for wear and replace as needed 8 Use your foot to press down on the foot switch to allow water to reac...

Page 21: ...he surface when treating sensitive areas Treat less sensitive area first and then return to sensitive areas If the problem persists reduce the power output intensity of the hand piece Section VIII System Maintenance Daily start up 1 Open the water supply valve 2 Turn the KRUUSE SP1 on using the main power switch Verify that the LED power indicator LED is lit 3 Set the power control knob to full po...

Page 22: ...t to the hand piece is insufficient the insert is probably worn out Inserts should typically be replaced after six months or when worn out E Please contact a local authorized agent for technical support Section IX Troubleshooting While Bonart offers repair service the following trouble shooting procedures can help you save time 9 A Troubleshooting Guide Basic procedures 1 Check all lines and conne...

Page 23: ...888 526 6278 or 626 600 5330 Monday through Friday 9 00 AM to 5 00 PM Pacific Time For other areas contact your local Bonart representative X Specifications Classification Protection against electric shock Class I Degree of protection against electric shock Type BF Mode of operation Continuous Medical device directive IIa Standards Followed ISO13485 Medical Devices Quality Management Systems Requi...

Page 24: ...5cm L x 22 5cm W x 8 5cm H Weight 4 Kg include hand piece Hand piece Cable 250 cm Footswitch Cable 250 cm Operation environment Temperature 0 40 Relative Humidity 10 90 non condensing Transport and storage conditions Temperature 0 60 Relative Humidity 10 90 non condensing Atmospheric pressure 860 1060 hPa Section XI Disposal Please follow county and state regulations for disposal of the KRUUSE SP1...