compensate  for  changes  in  sample  height.  The  Boekel  UV  Crosslinker 
may  be  used  in  either  Time  Mode,  in  which  the  unit  runs  for  a  user-
defined amount of time, or Energy Mode, in which the unit runs until a 
user-defined amount of energy has been accumulated.   

 

This unit comes equipped and calibrated with a quantity of (5) 254 nm 
Shortwave  bulbs.    Should  the  need  arise  for  use  of  a  different  UV 
wavelength,  the  bulbs  can  be  swapped  for  bulbs  of  a  different 
wavelength  and  the  new  wavelength  can  be  easily  selected  in  the 
Boekel  UV  Crosslinker  Setup  menu.  The  Boekel  UV  Crosslinker 
automatically applies calibration settings for the new bulb wavelength, 
maintaining  accurate  readings  for  sensed  UV  energy  in  both  Energy 
Mode  and  Time  Mode.  There  is  no  need  for  the  user  to  manually 
recalibrate the unit when changing bulbs. 

 

3.0 

Assembly 

 

Remove the packing materials carefully and retain for future shipment 
or  storage  of  the  unit.    Inspect  for  damage.    Report  all  shipping 
damage to the carrier immediately.  Shipping damage is covered by 
the  carrier  and  repair/replacement  for  shipping  damages  must  be 
coordinated  through  the  carrier.    Complete  and  return  the  Warranty 
Registration Card.  Each package should contain: 

 

1 UV Crosslinker 

 

1 Power Cord 

 

5 each 254 nm Shortwave Bulbs 

 

Operating Instructions 

 

Warranty Card 

 

4.0 

Identification of Controls - 

See Figure 1 and 2 on Page 6

 

1. 

LCD Display 

2. 

Power Key – used to toggle power to unit 

3. 

Mode  Key  –  used  to  toggle  between  Time  mode  and  Energy 
mode 

4. 

Reset Key – used to reset parameters after completion of a cycle 

5. 

Start/Stop Key – used to start and stop cycles 

6. 

Up/Down Arrow Keys – used to increase or decrease parameters 

7. 

AC Power Socket – connection point for power cord 

8. 

Height Adjustment Knob – used to adjust the height of the bulbs 
relative to the chamber floor 

Summary of Contents for 234100

Page 1: ...N2400340 Rev 2 Page 1 BOEKEL UV CROSSLINKER Model 234100 115VAC 175W 1 6A 234100 2 230VAC 175W 0 8A OPERATING INSTRUCTIONS...

Page 2: ...Identification of Controls 5 5 0 Installation 7 6 0 Operation 7 7 0 Technical Specifications 9 8 0 Maintenance and Service 9 8 1 Calibration 9 8 2 Changing Bulbs 10 8 3 Cleaning 11 8 4 Contrast Adjust...

Page 3: ...traviolets Prot ger les yeux et la peau des rayons ultraviolets nocifs Always observe the following safety precautions Avoid exposure to direct or strongly reflected germicidal ultraviolet rays Germic...

Page 4: ...h all applicable disposal laws Do not position the equipment in a manner that makes it difficult to disconnect the power cord from the receptacle on the back of the unit Protection provided by the equ...

Page 5: ...he unit when changing bulbs 3 0 Assembly Remove the packing materials carefully and retain for future shipment or storage of the unit Inspect for damage Report all shipping damage to the carrier immed...

Page 6: ...N2400340 Rev 2 Page 6 Figure 1 Front View Figure 2 Back View 1 LCD DISPLAY 2 POWER KEY 3 MODE KEY 4 RESET KEY 5 START STOP KEY 6 UP DOWN ARROW KEYS 7 AC POWER SOCKET 8 HEIGHT ADJUSTMENT KNOB...

Page 7: ...gy Mode is selected The Up and Down keys are used with a speedup algorithm to set the desired target energy value If the Mode key is pressed the colon for time blinks showing that Time Mode is selecte...

Page 8: ...ws the following for 1 second The user is returned to the main screen in paused mode where they can adjust the target energy or time CYCLE PAUSED blinks every 2 seconds to remind the user that the cyc...

Page 9: ...he requirements of CAN CSA C22 2 No 61010 1 second edition including Amendment 1 or a later version of the same standard incorporating the same level of testing requirements 8 0 Maintenance and Servic...

Page 10: ...cement of pins in sockets at both ends of each bulb that is installed 8 2 2 To use a different wavelength the entire set of 5 bulbs must be replaced with a set of 5 bulbs of the desired wavelength See...

Page 11: ...agents other than mild soap and water on the viewing window as damage to the window may result 8 4 Contrast Adjustment 8 4 1 To increase or decrease the contrast level of the display power the unit on...

Page 12: ...t bulbs are operational and replace blown bulbs as needed Duration of Energy Mode significantly longer than usual for given amount of energy a Blown bulb s a Ensure all light bulbs are operational and...

Page 13: ...rtificate must be completed signed by the user and returned to Boekel Scientific in order to receive the RMA number Please be sure to mark the outside of the returned goods package with the RMA number...

Reviews: