Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ZZABP01 is intended for use in the electromagnetic environment specified below. The
customer or the user of the ZZABP01 should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment–
Guidance
Electrostatic
discharge
(ESD)IEC
61000-4-2
±8 kV contact
±2 kV, ±4 kV , ±8 kV
±15 kV air
±8 kV contact
±2 kV, ±4 kV , ±8 kV
±15 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines 100
kHz repetition
frequency ±1 kV for
input/output lines
±2 kV for power
supply lines 100
kHz repetition
frequency ±1 kV for
input/output lines
Main power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±0.5 kV, ±1 kV
differential mode
line-line
±0.5 kV, ±1 kV
differential mode
line-line
Main power quality should be that
of a typical commercial or hospital
environment.
Voltage
dips, short
interruptions,
and voltage
variations on
power supply
input lines
IEC 61000-4-11
0% U
T
(>100 % dip in U
T
)
for 0.5 cycle at 0º,
45º, 90º, 135º,
180º, 225º, 270º,
and 315º
0% U
T
(100 % dip in U
T
) for
1 cycle 0º
70 % UT
(30 % dip in U
T
)
for 25/30 cycles
at 0º
0 % U
T
(>100 % dip
in U
T
) for 250/300
cycles at 0º
0% U
T
(>100 % dip in U
T
)
for 0.5 cycle at 0º,
45º, 90º, 135º,
180º, 225º, 270º,
and 315º
0% U
T
(100 % dip in U
T
) for
1 cycle 0º
70 % UT
(30 % dip in U
T
)
for 25/30 cycles
at 0º
0 % U
T
(>100 % dip
in U
T
) for 250/300
cycles at 0º
Main power quality should be
that of a typical commercial
or hospital environment. If a
dip or an interruption of main
power occurs, the current of the
ZZABP01 may be dropped off from
normal level. It may be necessary
to use uninterruptible power
supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m, 50/60H
30 A/m, 50/60Hz
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.