A. Instructions for use

1.  ZZABP01 needs special precautions regarding ElectroMagnetic Currents and to 

  be installed and put into service according to the EMC information provided in 

  the section "Accompanying Documents";
2.  Portable and mobile RF communications equipment can affect ZZABP01.

B. Technical description

1.  Warning that the use of accessories, transducers, and cables other than those 

  specified with the exception of transducers and cables sold by the manufacturer 

  of the ZZABP01 as replacement parts for internal components may 

  result in increased emissions or decreased immunity of the ZZABP01.
2.  Warning that the ZZABP01 should not be used adjacent to or 

  stacked with other equipment.

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions

The ZZABP01 is intended for use in the electromagnetic environment specified below. The 
customer or the user of the ZZABP01 should assure that it is used in such an environment.

Emissions

Compliance

Electromagnetic Environment–Guidance

RF emissions

CISPR 11

Group 1

The ZZABP01 uses RF energy only for its 

internal function. Therefore, its RF emissions 

are very low and are not likely to cause any 

interference in nearby electronic equipment.

RF emissions

CISPR 11

Class B

The ZZABP01 is suitable for use in all 

establishments, including domestic 

establishments and those directly connected 

to the public low voltage power supply network 

that supplies buildings used for domestic 

purposes.

Harmonic

emissions

IEC 61000-3-2

Class A

Voltage

fluctuations/

flicker

emissions

IEC 61000-3-3

Complies

Accompanying Documents

Summary of Contents for ZZABP01

Page 1: ...ZZABP01 READ THIS INSTRUCTION MANUAL CAREFULLY BEFORE USE Instruction Manual Digital Blood Pressure Monitor...

Page 2: ...r Replace Batteries 7 Time and Voice On Off System Setup 8 Unit Conversion Display mmHg kPa 8 WHO Blood Pressure Classification Display 9 Attaching the Arm Cuff 9 How to Take Proper Measurements 10 Me...

Page 3: ...to diabetes liver disease hardening of the arteries etc there may be fluctuation in blood pressure values measured at the upper arm versus at the wrist 3 Measurements may be impaired if this device i...

Page 4: ...on and should not perform self judgment and self treatment by the measured result Self diagnosis from measured results and subsequent treatment is dangerous and the device should not be used to judge...

Page 5: ...anual The use of other parts not approved by the manufacturer may cause faults or injuries For service information parts list or more information please contact BodyMed directly Do not submerge the de...

Page 6: ...effects of outside influences on blood pressure readings supplements the doctor s readings and provides a more accurate complete blood pressure history 4 WHO blood pressure classification Standards fo...

Page 7: ...ed collecting party subject to the regulation of each individual territory Insert the batteries as shown in the battery compartment If not the device will not work When LOW BATTERY mark blinks in the...

Page 8: ...SET key until the year number displays and flashes on the LCD to enter setting mode 2 Press MEM key to adjust the year then press SET key again to save your setting and enter the month setting mode 3...

Page 9: ...ause a faint pulse and result in a measurement error 2 Constriction of the upper arm caused by rolling up a shirt sleeve may also prevent accurate readings 3 Secure the arm cuff with Velcro strip so t...

Page 10: ...hin an hour after dinner after ingesting wine coffee or red tea after sports or physical activity after bathing after talking excessively or anxious behavior after moving around including bending forw...

Page 11: ...MEMORY To delete stored readings press and hold the MEM button for five seconds The LCD screen will display when memory has been deleted To keep your Digital Blood Pressure Monitor in the best conditi...

Page 12: ...rox 420g excluding batteries Classification Type BF Upper Arm Circumference 22 52 cm Specifications may be changed without notice in the event of improvements being made 1 Type of protection against e...

Page 13: ...ement Hand or body shaking during measurement Keep body still and measure again Battery icon on Battery has low power Replace battery and measure again The systolic pressure value or diastolic pressur...

Page 14: ...uipment Guidance and Manufacturer s Declaration Electromagnetic Emissions The ZZABP01 is intended for use in the electromagnetic environment specified below The customer or the user of the ZZABP01 sho...

Page 15: ...0 5 kV 1 kV differential mode line line Main power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input...

Page 16: ...and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 283 MHz and 40 66 MHz to 40 70 MHz The amateur radio bands between 0 15 MHz and 80 MHz are 1 8 MHz to 2 0 MHz 3 5 MHz t...

Page 17: ...hase is required before full replacement will be provided Limited Product Warranty For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m c...

Page 18: ...18 REORDER NO ZZABP01 Manufactured for BodyMed Hudson Ohio 44236 1 866 528 2152 Made in China Rev 030514 Digital Blood Pressure Monitor...

Reviews: