9.1.6
Power Supply technical data:
9.1.6.1
Characteristics:
Brand: GLOBTEK (HONG KONG) LTD
Frequency range: 50/60Hz
Input voltage: 100-240V
Output voltage: 5V
Input current: 0.2A
Output current: 1A
10
Failures guide
Tableau 3 : Failures guide
Problem
Probable cause
Solution
The monitor does not switch on.
Monitor in sleep mode
Check that the monitor is not in
sleep mode or deep sleep mode
The battery of the monitor is
discharged
Connect it on the mains supply
and wait few minutes before
interacting the monitor (on the
mains supply)
The monitor is in end of life
The manufacturing date is printed
on the label. The proper operation
of the monitor is warranted for
500 recharges cycles.
The monitor may require a
maintenance action
Return to your distributor or to the
manufacturer.
The LED of the activator does not
switch on.
The activator is not property
connected
Ensure that the connections are
correct and the power outlet has
power
The activator is in end of life
The manufacturing date is printed
on the label. The proper operation
of the activator is warranted for 2
years
The activator may require
maintenance action
Return to your distributor or to the
manufacturer
The RF communication between
monitor and capsule is not
working.
The distance is too large
Ensure that the capsule is in the
range of the monitor, check the
date of the last temperature data
received.
The capsule is not associated
Respect the activation process.
If the association is difficult,
ensure that the capsule is close
enough to the monitor or please
turn the capsule in the hole of the
activator.
The monitor indicates the number
of associated capsules.
Inappropriate autonomy of the
monitor
Non-recharged battery
Connect the monitor to the mains
supply and wait few minutes
before interacting with the monitor
(on the mains supply).
Battery in end of life
Scrap it to a DEEE organism.
Inappropriate autonomy of the
capsule
Old battery
Check the date printed on the
label.
Incorrect connection
Check that the cable is properly
connected.