User Guide e-CELSIUS Performance.2
EN
28
After an event “
monitor switch off/extinction
”, it needs to put the monitor in charge and to
set a correct date and time. After validation of the date and time on the monitor, the radio
communication with the capsules will get back and the data synchronization will begin. To
indicate this event, a ma
rker “monitor wake-up” will be stored and displayed on the download
window and the CSV file.
7.1.4
Hide markers
If you do not want to see all the markers, you can check the "Hide Markers" at the bottom left
of
Picture 27: Screen of marker management
7.1.5
Export
temperature’s data
unloaded on e-Performance Manager
software
To export temperature
’s data, one curve or more may be selected from the icons at the top
right of the screen, shown in
Picture 26: Picture of the curves obtained on the e-performance
Manager software after the data unloading.
To export curves as displayed on the screen of the software, a PDF file may be generated with
the "PDF" button (Fig. 26). The view from the graph in the exported PDF file is the display of
the graph (with the same zoom and the same number of curve).
A data file in spreadsheet format can be generated from the "Export" button. A spreadsheet
including the temperature data, the date and time of recording and the markers will be
generated automatically.
In the CSV file you will find the complete data table as shown on picture 28.
The status column gives important information:
Case 1: you have a temperature in column D and void status
: This data was received by
the monitor in real time.
Case 2: you have a temperature in column D and Synchronised status
: this data was
asked by the monitor afterwards during synchronisation process.
Case 3: you have no data in column D and you have a “out of limit” status
: this data was
not synchronised yet OR this data is out of the range of synchronisation number set on the
monitor.