User Guide e-CELSIUS Performance.2
EN
20
If the LED of the activator stops flashing and the message "Error! Would you
like to restart an activation " appears on the monitor screen, please check the
positioning of the e-Celsius
®
performance capsule in the hole of the Activator
and / or slightly move the e-Celsius
®
performance capsule in the hole, press the
button OK on the monitor to restart the association process and re-press then
the button on the Activator. It may be wise to rotate the pill inside the hole during
activation progression in order to optimize the process.
Be careful: It is highly recommended to verify the environment table used
for activation. If the environment is metallic, it could be difficult or even
impossible to activate pills properly.
For the activation of an additional capsule, repeat the procedure. It is possible to connect up
to 3 capsules in parallel with a single monitor.
6.7.4
Consult temperature data in real time
In order to visualize the collected temperature data, go to the menu "DATA" and the submenu
"data view" or directly by pressing the button Home
. The screen will then allow you to
visualize the latest temperature data collected for each activated capsule and the minimum
and maximum values collected by each associated capsule.
Temperature data (real time temperature data, min and max) of 1 to 3 capsules may be
displayed on the screen.
When this visualization mode in used, the monitor screen never turns off to standby
mode.
Please note that, when the visualization mode is used, the data synchronization
between the monitor and the capsules is slowed. It is advisable to leave this screen to
optimize synchronization delay.
Picture 19: Home screen menu of data view