3.4

Elec tri cal con nec tions

3.4.1

In tro duc tion

Use a suitable connector mating half (not supplied) to connect the electrical
supplies and your control equipment to the connector on the logic interface cable
(Figure 1, item 3).  When you make the electrical connections to the EXDC
described in the following sections, refer to Table 1 for full details of the logic
interface connections and to Figure 3 for a schematic diagram of the connections.

Pin number Signal

Polarity Use

1

Start/Stop control input

-

Close * to start EXT
pump.

2

TMP Normal status output

-

Closed * when EXT
pump speed is 80%
or more of full
rotational speed.

3

Earth (ground)

-

4

Pump speed analogue output

Positive

5, 6

Electrical supply: 0 V

-

7, 8

Electrical supply: 80 V

Positive

 * With respect to pins 5 or 6.

Table 1 - Logic interface connector pins

3.4.2

Con nect the logic in ter face to your con trol equip ment

1. Connect your control equipment to the start/stop control input pins of your

logic interface mating half to start/stop the EXT pump as required:

Link (close) the start/stop con trol in put (pin 1) to pins 5 or 6 to start the
EXT pump.

Un link (open) the start/stop con trol in put (pin 1) from pins 5 or 6 to stop
the EXT pump.

2. Connect your control equipment to the pump speed analogue output (pin 4)

and to pins 5 or 6 of your logic interface mating half to monitor the speed
output of the EXDC.

(Continued on page 10)

8

EXDC Tur bo mo lecu lar Pump Drive Mod ules

Summary of Contents for D396-40-000

Page 1: ...UK Telephone 44 0 1293 528844 Fax 44 0 1293 533453 http www bocedwards com Instruction Manual EXDC Turbomolecular Pump Drive Modules Description Item Number EXDC80 Turbomolecular Pump Drive Module D39...

Page 2: ......

Page 3: ...EXT pump 7 3 4 Electrical connections 8 3 4 1 Introduction 8 3 4 2 Connect the logic interface to your control equipment 8 3 4 3 Connect the electrical supply 10 4 OPERATION 11 4 1 Start up 11 4 2 Ope...

Page 4: ...ED fault finding 14 Associated Publications Publication Title Publication Number EXT Pump Accessories B580 66 880 EXT70 and EXT250 Turomolecular Pumps B722 01 880 EXT351 and EXT501 Turbomolecular Pump...

Page 5: ...tion is highlighted as WARNING and CAUTION instructions you must obey these instructions The use of WARNINGS and CAUTIONS is defined below WARNING Warnings are given where failure to observe the instr...

Page 6: ...e end of the logic interface cable Figure 1 item 3 You must use a suitable connector mating half not supplied to connect the EXDC to your own equipment refer to Section 2 3 for the connector mating ha...

Page 7: ...ecular Pump Drive Modules 3 Figure 1 The EXDC 1 Grub screw 2 Cooling fins 3 Logic interface cable 4 Phase LEDs red 5 TMP Normal LED green 6 Connector lock 7 Unlocked position 8 Locked position 9 EXT p...

Page 8: ...50082 1 2 2 Mechanical data Dimensions See Figure 2 Mass 0 2 kg Enclosure protection IP20 as defined by IEC529 Pollution degree IEC664 category 1 2 3 Electrical data Electrical supply requirements The...

Page 9: ...EXDC Turbomolecular Pump Drive Modules 5 Figure 2 Dimensions mm A Cable bend radius 30...

Page 10: ...d Maximum input current 1 1 A EXDC80 2 2 A EXDC160 Start stop control input signal Start control voltage low close 0 8 V d c Stop control voltage high open 4 to 24 V d c Speed analogue output Output v...

Page 11: ...XDC Refer to Figure 1 Use the following procedure to adjust the position of the EXT pump connector 9 and connector lock 6 with respect to the EXDC body 1 Use a hexagonal key to loosen the grub screw 1...

Page 12: ...ational speed 3 Earth ground 4 Pump speed analogue output Positive 5 6 Electrical supply 0 V 7 8 Electrical supply 80 V Positive With respect to pins 5 or 6 Table 1 Logic interface connector pins 3 4...

Page 13: ...ic diagram of the logic interface connections A EXDC logic interface plug B Mating half to EXDC logic interface plug 4 D C relay optional 5 Relay power supply optional 6 Voltmeter 1 Start stop switch...

Page 14: ...ce in the electrical supply close to the EXDC If you do not you will not be able to switch off the EXDC in an emergency WARNING Incorporate suitable fuses or current limiting devices in the electrical...

Page 15: ...EXT pump may not reach 80 of full rotational speed and the TMP Normal status output signal will not be set Use the following procedure to start up your system This procedure assumes that you will man...

Page 16: ...maximum allowable vent rate 1 Select Stop open the Start Stop control input on the logic interface connector 2 Open the vent valve after the EXT pump speed has fallen to below 50 of full rotational s...

Page 17: ...to Start the EXDC will automatically restart the EXT pump when the electrical supply is restored after an electrical supply failure Ensure that people cannot be injured by the rotating rotor blades of...

Page 18: ...stem Inspect the vacuum system and seal any leaks found Is the EXT pump faulty Inspect the EXT pump and check that it operates correctly with another EXDC or EXC Turbomolecular Pump Controller If nece...

Page 19: ...clean dry conditions until required When required for use prepare and install the EXDC as described in Section 3 of this manual 6 2 Disposal Dispose of the EXDC and any components safely in accordanc...

Page 20: ...hazardous materials Note Some contaminated equipment may not be suitable for airfreight PROCEDURE 1 Contact BOC Edwards and obtain a Return Authorisation Number for your equipment 2 Complete the Retur...

Page 21: ...roof of decontamination YOU MUST CONTACT BOC EDWARDS FOR ADVICE BEFORE YOU RETURN SUCH EQUIPMENT Print your name _________________________________Print your job title _________________________ Print y...

Page 22: ...7KLV SDJH LQWHQWLRQDOO EODQN...

Page 23: ...631 3501 Fax 1 514 631 3502 CHINA BOC TRADING SHANGHAI CO LTD 23 Fu Te Road N Wai Gao Qiao Free Trade Zone Pudong Shanghai 200131 PRC China Tel 86 21 58669618 Fax 86 21 58669993 GERMANY BOC EDWARDS Am...

Reviews: