23
3.8.2 Gun
"Ready"
When all the interlocks are satisfied, "Closed", the circuit is ready for the gun to be turned ON.
The indicator in the gun off switch, G1-I2, will be illuminated. The circuits that must be
satisfied for gun operation are:
Remote
Off
G1-J2
Green-Black
Gun
Control
Off
G1-S1
Control
Panel
Tank
Interlock
G1-J2
Orange-Black
Vacuum
Interlock G1-J2
Green-Red
Auxiliary
Interlock G1-J2
Blue-Red
Water
Interlock
G1-J2
Yellow-Red
Position Supply Adjustment
G1-J2 Brown-Red
3.8.3 Gun
"ON"
The "Gun ON" command turns on the filament power supply. The filament supply is
controlled by a closed loop regulated circuit. At turn ON, the filament ramps up and regulates
to the current level commanded by the bias potentiometer, a screwdriver adjustment on the
control panel. If high voltage is ON, the high voltage regulator sends a signal to the filament
regulator to transfer the reference command from the bias adjustment potentiometer to the
emission current potentiometer, also located on the control panel, with a calibrated 10 turn dial.
If high voltage is off, the filament current command adjustment stays with the bias adjustment
pot until such time the high voltage is turned ON.
The output vacuum relay, located in the "filament transformer module", switches the high
voltage output onto one leg of the high current filament transformer, secondary. In a single gun
system, the relay does not transfer until both the high voltage and gun circuits are turned ON.
3.8.4
Gun Power Supply Metering
The 3 1/2 digit LCD meter has a selector switch. G1-S4 that provides the operator with three
circuit-monitoring positions.
Emission current: The current is monitored in the high voltage leg of the gun circuit. The
maximum emission current is read as "600"mA.
Filament Current: The primary current of the filament transformer is monitored, rectified and
provided as an external readout (7.5V = 75 amperes) and the meter will read "75.0" amperes.
A remote filament current monitor is provided as 7.5V = 75 Amps.
Position Current: The output current of the position power supply is monitored and will read
"3.00" amperes. There is no external readout for position current.
Summary of Contents for CV-6SL
Page 1: ...1 CV 6SL INSTRUCTION MANUAL 6000W POWER SUPPLY...
Page 2: ......
Page 5: ...5...
Page 12: ...12 Sample Remote I O Connections...
Page 15: ...15...
Page 21: ...21...
Page 28: ...28...
Page 30: ...30...
Page 44: ......
Page 45: ......
Page 46: ......
Page 47: ...1 F E C D B A 1 2 3 4 5 6 2 3 4 5 6 7 8 F E 7 8 C D B A D...
Page 48: ...1 F E C D B A 1 2 3 4 5 6 2 3 4 5 6 7 8 F E 7 8 C D B A D...
Page 49: ......
Page 50: ......
Page 51: ......
Page 52: ......
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ......
Page 57: ......
Page 58: ......
Page 59: ......
Page 60: ......
Page 61: ......
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ...1 F E C D B A 1 2 3 4 5 6 2 3 4 5 6 7 8 F E 7 8 C D B A D...
Page 67: ...1 F E C D B A 1 2 3 4 5 6 2 3 4 5 6 7 8 F E 7 8 C D B A D...
Page 68: ......
Page 69: ......
Page 70: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 84: ...51...