OZONE MONITOR BMT 932 Manual, Rev. 02/2021
50
Warning:
This product relies on the building's installation for short-circuit (overcurrent) protec-
tion. Ensure that a fuse or circuit breaker no larger than 15 A at 120 VAC (10 A at 240 VAC) is
used on the phase conductor.
Warning:
Disconnect electrical power before opening the cabinet door.
Précaution
: Couper l'alimentation avant d'ouvrier la porte d'analyseur.
Signal contacts (Connector "Output & Control") are as described on page 15. A splash-proof
connector for connection of the supplied serial cable is provided.
The BMT 932 C can be ordered with the option 932C-HAR. This provides a relay which can
switch an alarm horn in case of a High Concentration Alarm. Connection is made through an
additional splash-proof feed-through to the two pole connector on the PCB labeled "Horn". Ca-
ble diameter shall be 6.5 to 10 mm.
Warning:
The equipment connected to the option 932C-HAR shall be fused at 4 A.
Précaution
: Assurer la protection de l’équipement connecté au option 932C-HAR avec une fusible pour 4A.
C4
Operation
The operation of the BMT 932 C is the same as of the standard BMT 932.
For further description of the functions and properties of the cabinet version please refer to the
main part of this manual.