OZONE MONITOR BMT 932 Manual, Rev. 02/2021
28
Testing Scrubber
Please wait
After test completion the result is displayed:
Testing Scrubber
Scrubber is OK
If a faulty scrubber was found there will be a Scrubber Alarm.
See p. 10 and 34 for a more detailed desripton of scrubber operation.
Time/Date
Time
Here the time of day is set in the format hh:mm:ss.
Date Format
Display of date can be switched from European (DD.MM.YY) to American
(MM/DD/YY) notation.
Date
During entry the date will be constantly tested on conformance to the calender
rules, so it is not possible e.g. to enter the date 29.02.01. In order to make this test
possible, first the year, then the month and then the day has to be set.
Other Parameters
Set Active Channels
This item can be used to select the channels which shall be sampled. Non-active
channels are skipped in Auto mode and are not selectable in Manual mode. Con-
centration alarm setups and EMO assignment are only possible for active chan-
nels. The display reads:
Set Active Channels
1Y 2Y 3Y 4Y 5N 6Y
In this example all channels except channel 5 are active.
Set Sniffer On / Off
Sniffer mode can be activated here. When entering this mode, the instrument
completes one last zero cycle and then displays concentrations in 1 s intervals.
This is useful for finding ozone leaks by using a PTFE tube (with sample gas filter
attached) and "sniffing” in the area of an installation where a leak is suspected.
Channels can still be changed in Sniffer mode. A channel change will lead again to
one single zero cycle. Because the instrument will show a zero drift in this mode,