BMS EASY-CHECK FN Operational Manual Download Page 7

Operation 

EASY- CHECK FN 

Manual 

    Rev:00                                                           

 7  / 10

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

General Remarks 

The probe should 

not be drawn across the testing surface

 but reset at dif-ferent 

spots, i.e. 

after each measurement hold the instrument in the air for about 1 sec

In doing so the stored calibration is automatically checked and corrected if nec-
essary. 

Make sure that the probe surface and the base plate are kept clean and  
polished at all times.  

 

When measuring on small or curved objects it is advisable to perform  
calibration on a bare test object with the same geometry of the object to be meas-
ured and not

 

on the base plate supplied with the device. 

The device switches off automatically one minute after the last measurement. The 
instrument can also be switched off with the red key. 

Thickness of the base material:

 

Base material steel FE:     

> 300 µm 

Base material non-ferrous metal NFE:  >   50 µm 

Exchange Of The Battery  

As soon as the symbol 

<BAT> 

is flashing the battery must be exchanged by a new one.  

 

When the voltage of the battery is less than 0.8 V the device switches off by itself. 

 

Please insert only leak proved batteries ! 

 

Charge The 1.2v Rechargeable Battery With The Line-Charging Unit 

 

As soon as the symbol 

<BAT> 

is flashing the rechargeable 1.2V Mignon battery must be 

charged.  

Important:

 Before inserting the charging connector into the interface plug of EASY-

CHECK FN the device must be switched off. 

The charger must not be connected when the 1.5V battery is inside the device. The 
battery may run out and destroy the device ! 

With the device switched off the charging connector is inserted into the interface plug of 
EASY-CHECK FN. The device switches on by itself and a bar diagram with the battery 
symbol is displayed. With the charger connected you can go on taking measurements. 
After approx. 20 sec the state of charge is displayed again provided no measurement is 
taken. When the charger is connected the automatic switch off is deactivated. 
The rechargeable battery needs approx. 8 hours to be charged completely. A constant bar 
diagram is flashing together with the battery symbol and the charger can be removed. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for EASY-CHECK FN

Page 1: ...NAL MANUAL Bms Bulut Makina Sanayi Ve Ticaret Ltd Şti İkitelli Organize Sanayi Bölgesi Dolapdere Sanayi Sitesi Ada 4 No 7 9 İkitelli İSTANBUL TÜRKİYE Phone 90 212 671 02 24 671 02 25 Fax 90 212 671 02 26 web www bulutmak com e mail bms bulutmak com ...

Page 2: ...IcL Clear the last reading 5 cLr Clear the whole Data Memory 5 nFE Base material recognition 5 Un Change unit of measurement µm mils 5 rES Resolution 6 Calibration 6 General Remarks 7 Exchange Of The Battery 7 Charge The 1 2v Rechargeable Battery With The Line Charging Unit 7 Technical Data 8 Optional Accessories 8 Software Incl Operation Instructions 8 Radio Control 9 Data Printer Mega Print 9 Da...

Page 3: ...ed steel without considering the zinc thickness In addition EASY CHECK FN includes the following features Store up to 1000 readings in the data memory Display statistics Transfer all readings and statistics to a printer or computer by means of the interface or just Transfer the data by RADIO All these functions can be performed with the red key And This Is Really Unique The device recognizes by it...

Page 4: ...Modes Sto Store readings in the Data Memory To store readings in the memory for further evaluation press the key until Sto is dis played Now release the key and wait until on is flashing Confirm by pressing the key again The data memory is activated and all subsequent measurements will be stored The sym bol STORE is displayed to make evident that the memory is switched on Capacity of the data memo...

Page 5: ...nt and the zinc coating together because it is not able to recognize the thin zinc coating as non ferrous base ma terial nFE To measure only the paint coating on the zinc plating the device must be switched over to the nFE mode To avoid wrong measurements the following requirements should be met The zinc plating should have a thickness of at least 20 µm The zero adjustment should be performed on a...

Page 6: ...e device on 2 Set the device with the probe on the base plate FE blue and press the key for about 2 sec until 0 0 is displayed stable Then release the key and zero setting is per formed and confirmed by a beep signal 3 Take off the device the foil value entered previously is flashing 4 To enter the value of the shim press the key as long as the value scrolls up Pressing the key shortly the value g...

Page 7: ...nged by a new one When the voltage of the battery is less than 0 8 V the device switches off by itself Please insert only leak proved batteries Charge The 1 2v Rechargeable Battery With The Line Charging Unit As soon as the symbol BAT is flashing the rechargeable 1 2V Mignon battery must be charged Important Before inserting the charging connector into the interface plug of EASY CHECK FN the devic...

Page 8: ... MEAN STD DEV Power Supply 1 5V Mignon battery 1 2V rechargeable battery with charger available Recording Data on steel FE one long beep on non ferrous metals NFE two short beeps Measuring Probe swinging by 90 Dimensions 108 x 48 x 38 mm Weight approx 100 g Interface serial RS 232 C 5 V TTL level Baudrate Printer PC 1200 baud Data Stop bits Printer PC 7 2 Warranty Indication unit 12 months Measuri...

Page 9: ...t in NiCad battery must be charged The built in NiCad battery is charged with the charging unit supplied with the printer The cable of the charging unit is connected at the right hand socket of MEGA PRINT The charging time should be at least 4 hours Operating Instructions 1 The operation of MEGA PRINT together with the Coating Thickness Meter EASY CHECK FN is explained in the operating instruction...

Page 10: ...Insert a new paper roll Open the lid Insert the paper roll Pull out the end of the paper Close the lid Operation EASY CHECK FN Manual Rev 00 10 10 ...

Reviews: