Hydrosplit M3 Manuale utente – User manual - v1.6.5_ITA_ENG
56
CALCULATION UNIT DATA SHEET
Model
Hydrosplit M3
Power supply
-Battery supply
-Electrical supply (optional, 3.6 ÷ 5 VDC, 300 mA): in this case, the battery
will be used as backup power supply)
Battery type
Li-SoCl2 Lithium-thionyl chloride, 3,6V “size D” 20Ah
Battery Life
10 years +1
Operating temperature range
5 ÷ 55°C
Storage temperature range
-10 ÷ 55°C
Dimensions
154 x 119 x 51mm
IP protection grade
IP52
Approval
2004/22/CE MID (Module B)
Certificate nr.IT-MI004-xxxx
EN1434
Environmental class
A (E1, M1)
Temperature measurement range, MID
approved (heating applications)
Θ: 5÷180°C
Temperature difference range, MID
approved (heating applications)
ΔΘ: 3÷150K
Temperature measurement range
(cooling)
Θ: 2÷24°C
Temperature difference range
(cooling)
ΔΘ: 3÷20K