IntraOs 70 – Service & Installation M
30/52
69 500 00210
ARM FIRST
SECTION
MINIMUM
CLEARANCE
ARM SECOND
SECTION
a) BASIC PIN
b) BRASS BUSHING
c) SAFETY CLA
d) SIDE HOLDER
PARTS OF ARTICULATED JOINT
a
b
c
d
4.3
Maintenance of the Folding Arm
Specifically for the folding arm inspect for wear of pins and levers
every 24 months, in case of normal use (about 6000 cycles). Replace the
worn out parts or the arm itself if it appears damaged.
Maintenance can be performed with the arm mounted on the extension arm
or on the mobile stand.
A
Disconnect the tube housing assembly keeping the arm fully open to
avoid the risk of injuries (fig.1) .
B
Remove the snap-out caps of the first arm section, the blocking screws
of the pins, then the metal cover by pushing the pins out paying
attention to supporting the arm (fig. 2 and fig. 3).
Service & Installation Manual
Version 4.0
a) BASIC PIN
b) BRASS BUSHING
c) SAFETY CLAMP
d) SIDE HOLDER
PARTS OF ARTICULATED JOINT
inspect for wear of pins and levers at least
e (about 6000 cycles). Replace the
Maintenance can be performed with the arm mounted on the extension arm
bly keeping the arm fully open to
caps of the first arm section, the blocking screws
of the pins, then the metal cover by pushing the pins out paying