3.1.15 Final Tuning and Set-Up
A
Remove the line voltage fuse(s) from the timer
B
Put back the cover plate on the folding/simple arm (tube-
C
Tighten locking screw and tune friction screw in Extension Arm.
D
Tune friction screw on the folding arm for rotation around the vertical
axis.
E
Make sure the Timer switch is in the “off” position.
F
Set the dip switches SW2 as indicated in the following to define:
a
Enable/disable correction of exposure time to compensate
fluctuations of line voltage.
b
Indicate the nominal line voltage of the tube-head (see label).
c
Set the pre-heating time according to type of X-ray insert
tube-head assembly; the measured value is marked in front of the
tube-head;
G
Put back the line voltage
fuse(s).
H
Put back all plastic covers
and logo strips.
I
Connect the system to the
line voltage by switching the
general line voltage switch
on and/or plug the power
cord into the wall socket.
J
Switch the timer “on”.
K
The system is now ready for
Functional Check as
described in the Operator’s Manual.
PRE-HEATING TIME SETTING - TYPICAL VALUES
Type of insert
OCX/70-G
RF8G070
Pre-Heating
Time
180 ms
160 ms