IntraOs 70 – Service & Installation Manual
Version 4.0
69 500 00210
3.2
Mobile Systems
3.2.1 Unpacking
Unpack the components of the system and
check the following:
A
Each item is in good conditions and was
not damaged during transportation.
B
All the items for the desired system
configuration are available.
C
The line voltage on the labels of timer
and tube-head are corresponding to the
existing local line voltage.
3.2.2 Room Preparation
Make sure that a wall socket is available
close to the dental chair in reach of the
mobile X-ray equipment to be connected.
Such a socket has to be provided with a
ground connector to the protective earth
point of the room. For the line voltage
supply to the wall socket it is recommended
the use of a cable with large section
conductors to keep to minimum the cable
resistance and the relevant voltage drop.
Wire Type
Square Section mm
2
Resistance Ohm/m
AWG 16
1.35
AWG 14
2.11
2x0.0094
25/52
Resistance Ohm/m
2x0.015
2x0.0094