12.0 Settings
11.1 Tube Calibration
This feature allows the user to calibrate the pump’s indicated flow rate to the system
To Calibrate Your Tube:
1
On the home screen select the
Calibration Icon
2
Enter values:
!
Pump Speed (RPM)
!
Run Time (seconds)
3
Select
Start
to begin
4
Select
Start
5
Enter the measured flow rate
into the field
6
Confirm by selecting
Save
®
FLEXFLO M3
11.0 Tubing
Page 36
To Input Pump Name:
1
Open the
App Drawer
2
Open
Settings
3
Open
System
4
Select “
Pump Name
”
5
Enter desired
Pump Name
6
Confirm by pressing
OK
12.1 Pump Name
This is to change the name of the pump that is displayed on the home screen.