
5.4
Wiring Terminals and I/O Schematics
Page 13
®
FLEXFLO M3
Risk of electric shock - All
wiring must be insulated
and rated 300V minimum.
!
WARNING
Shielded cables should be used on all input signal wires.
INPUT:
4-20 mA
FUNCTION
PIN #
RATING
BLOCK DIAGRAM
(+) POSITIVE
1
2
(-) NEGATIVE
(-)
(+)
ACTIVE 4-20mA
TRANSMITTER
SOURCE
(OPTIONAL PUMP #2)
Single or dual pump (series)
input. Loop voltage must not
exceed 24 Volts.
P4
P1
P2
P3
P5
P4
P1
P2
P3
P5
M12 Connector
INPUT #2
INPUT:
FVS SYSTEM
(FLOW VERIFICATION
SENSOR)
FV SENSOR ONLY
(+) POSITIVE
2
3
(-) NEGATIVE
P4
P1
P2
P3
P5
INPUT #1
RED (+)
BARE
BLACK (-)
BLUE-WHITE
FVS SENSOR
4
SIGNAL
INPUT:
FREQUENCY, AC
SINE WAVE, TTL,
CMOS
(+) POSITIVE
4
5
(-) NEGATIVE
(-)
(+)
FREQUENCY TRANSMITTER
SOURCE
P4
P1
P2
P3
P5
INPUT #2
INPUT:
FREQUENCY, AC
SINE WAVE, TTL,
CMOS
INPUT #1
BLUE-WHITE
MICRO-FLO
FLOWMETER
PULSE OUTPUT
P4
P1
P2
P3
P5
SIGNAL
BLACK (-)
(+) POSITIVE
2
3
(-) NEGATIVE
4
SIGNAL
INPUT:
REMOTE
START/STOP
DRY CONTACT C
PRIMARY
(+) POSITIVE
1
5
(-) NEGATIVE
P4
P1
P2
P3
P5
INPUT #1
(+)
(-)
OPEN CIRCUIT
IMPEDANCE MUST
BE GREATER THAN
50K OHM
OUTPUT:
4-20 mA
(+) POSITIVE
1
2
(-) NEGATIVE
(-)
(+)
P4
P1
P2
P3
P5
OUTPUT #1
4-20mA RECEIVER
600 OHM LOAD MAX.
DIGITAL PULSE
RECEIVER CIRCUIT
1.5K OHM
EXTERNAL SOURCE
6-30V DC
OUTPUT:
FREQUENCY-
OPEN COLLECTOR
(+) POSITIVE
1
2
(-) NEGATIVE
(-)
(+)
P4
P1
P2
P3
P5
OUTPUT #4
OUTPUT:
CONTACT
CLOSURE
#1
NORMALLY
OPEN
3
4
P4
P1
P2
P3
P5
OUTPUT #1
NC
C
NO
5
COMMON
(GROUND)
SWITCH LOAD
1 AMP MAX @ 125V AC
0.8 AMP MAX @ 30V DC
NORMALLY
CLOSED
OUTPUT:
CONTACT
CLOSURE
#2
NORMALLY
OPEN
3
4
P4
P1
P2
P3
P5
OUTPUT #2
NC
C
NO
5
COMMON
(GROUND)
SWITCH LOAD
1 AMP MAX @ 125V AC
0.8 AMP MAX @ 30V DC
NORMALLY
CLOSED
OUTPUT:
CONTACT
CLOSURE
#3
NORMALLY
OPEN
3
4
P4
P1
P2
P3
P5
OUTPUT #3
NC
C
NO
5
COMMON
(GROUND)
SWITCH LOAD
1 AMP MAX @ 125V AC
0.8 AMP MAX @ 30V DC
NORMALLY
CLOSED
OUTPUT:
RELAY
6 AMP
NORMALLY
OPEN
3
4
P4
P1
P2
P3
P5
OUTPUT #4
NC
C
NO
5
COMMON
(GROUND)
NORMALLY
CLOSED
SWITCH LOAD
6 AMP MAX @ 250V AC
5 AMP MAX @ 30V DC
INPUT:
AUTO-PRIME/
DRY CONTACT C
SECONDARY
(+) POSITIVE
1
5
(-) NEGATIVE
P4
P1
P2
P3
P5
INPUT #2
(+)
(-)
OPEN CIRCUIT
IMPEDANCE MUST
BE GREATER THAN
50K OHM
5.0
Installation
P4
P1
P2
P3
P5
5.3
Input Power Connections
Risk of electric shock – cord connected models are supplied with a grounding
conductor and grounding-type attachment plug. To reduce risk of electric shock, be
certain that it is connected only to a properly grounded, grounding-type receptacle.
WARNING
!
Be certain to connect pump to proper supply voltage. Using incorrect voltage will damage
pump and may result in injury. Voltage requirement is printed on pump serial label.
!
Input power range is 96VAC to 264VAC 50/60 Hz.
!
Voltage Selection is automatically detected and adjusted by power supply. No mechanical
switch necessary.
!
Use voltage your power cord is rated for.
!
Power cord models are supplied with a ground wire conductor and a grounding type attach-
ment plug (power cord). To reduce risk of electric shock, be certain that power cord is con-
nected only to a properly grounded, grounding type receptacle.
!
Be sure all M12 wiring cable glands are properly installed and sealed.
!
Never strap control (input / output) cables and power cables together.
!
Power Interruption:
This pump has a user programmable auto-restart feature which will can
either restore the pump to the operating state it was in when power was lost or require a user
action to restart.
Note:
When in doubt regarding your electrical installation, contact a licensed electrician.
Electrical connections and grounding (earthing) must conform to local wiring codes.
WARNING
Risk of electric shock - Disconnect electricity before removing the wiring
compartment cover.
WARNING
Page 12
®
FLEXFLO M3
5.0
Installation
P4
P1
P2
P3
P5
P4
P1
P2
P3
P5
P4
P1
P2
P3
P5
P4
P1
P2
P3
P5
P4
P1
P2
P3
P5
P4
P1
P2
P3
P5
P4
P1
P2
P3
P5
P4
P1
P2
P3
P5
P4
P1
P2
P3
P5
Summary of Contents for FLEXFLO M3 Series
Page 1: ...Peristaltic Metering Pump Series M3...
Page 6: ...2 0 Engineering Specifications Page 6 FLEXFLO M3...
Page 52: ...FLEXFLO M3 15 0 Replacement Parts List Page 52...
Page 53: ...Page 53 FLEXFLO M3 16 0 Tubing Information...
Page 56: ...FLEXFLO M3 Page 56 PAGE INTENTIONALLY LEFT BLANK...
Page 59: ...Page 59 FLEXFLO M3...