48
First Line and Planned Maintenance Manual
Frontline Sirius®
10. Check that an alternative means to ventilate the patient is immediately
available. (e.g. self-inflating bag and oxygen cylinder).
•
Check that the self-inflating bag and cylinder of oxygen are functioning correctly and the
cylinder contains an adequate supply of oxygen.
11. Recording
•
Sign and date the logbook kept with the anaesthetic machine to conform the machine
has been checked.
Record on each patient’s anaesthetic chart that the anaesthetic machine breathing
system and monitoring has been checked.