34
First Line and Planned Maintenance Manual
Frontline Sirius®
Slide onto mounting rods
To Scavenging
To CGO
To ventilator
Patient Circuit
TUBING ONLY APPLIES
TO SIRIUS 1000 & 2000.
2.7 Absorber Pre-Use Check
WARNING: Do not use a faulty absorber - injury to the patient may result. Call
an authorised service engineer or return the unit to the supplier.
1. Carry out a careful visual inspection of the absorber.
2. Ensure the absorbent canisters are full and installed. Ensure that all canister seals are
present and are correctly installed. (See Fig. 5).
3. Undo the two valve covers and check that both the inspiratory and expiratory valve discs
are clean and free to operate. Check the valve seats for chips, damage, etc. Replace the
discs and covers.
4. Ensure that the bypass control is in the ABSORBER ON position.
5. Ensure that the oxygen sensor port either has a sensor fitted or that the blanking plug is
fitted.
6. Ensure all tubing is correctly and securely installed.
Figure 4 Attaching Absorber Tubing
2. User Pre Use Testing
Summary of Contents for Sirius 1000
Page 17: ...16 First Line and Planned Maintenance Manual Frontline Sirius Notes ...
Page 27: ...26 First Line and Planned Maintenance Manual Frontline Sirius Notes 1 Technical Description ...
Page 37: ...36 First Line and Planned Maintenance Manual Frontline Sirius Notes 2 User Pre Use Testing ...
Page 61: ...60 First Line and Planned Maintenance Manual Frontline Sirius Notes 3 Full System Testing ...
Page 62: ...61 First Line and Planned Maintenance Manual Frontline Sirius 4 0 Maintenance 4 Maintenance ...
Page 71: ...70 First Line and Planned Maintenance Manual Frontline Sirius Notes 5 Routine Maintenance ...
Page 106: ...105 First Line and Planned Maintenance Manual Frontline Sirius 8 0 Spare Parts 8 Spare Parts ...