6
5
Maintenance
Visually check the device regularly.
Report any changes in performance of this device to the practitioner/service provider e.g.
unusual noises, increased stiffness, changes in heel cushioning, restricted/excessive movement,
significant wear or excessive discoloration from long term exposure to UV.
Inform the practitioner/service provider of any changes in body weight and/or activity level.
Cleaning
Use a damp cloth and mild soap to clean outside surfaces. DO NOT use aggressive cleansers.
The remaining instructions in this section are for practitioner use only.
This maintenance must be carried out only by competent personnel
(practitioner or suitable
trained technician).
The following routine maintenance is to be carried out at least annually:
• Check for damage or wear and replace if necessary.
Ensure the user has read and understood all safety and user-level maintenance information.
Advise the user that a regular visual check of the device is recommended and signs of wear that
may affect function should be reported to their service provider (e.g. significant wear or excessive
discoloration from long term exposure to UV).
Advise the user to inform the practitioner/service provider of any changes in body weight and/or
activity level
If this device is used excessively, the maintenance level and interval should be reviewed and if
required advice and technical support sought to plan a new maintenance schedule dependent
upon the frequency and nature of the activity. This should be determined by a local risk
assessment carried out by a suitably qualified individual.
4
Function
The device comprises a lightweight thermoplastic keel which gives a gentle rolling action
through stance. The keel lies within a standard closed foot shell profile and features a sandal toe
for a more natural appearance.
An adjuster screw in the heel allows the user to customize the hardness of the heel cushion for
individual requirements.
938086PK1/1-0121
Summary of Contents for P529101
Page 101: ...101 938086PK1 1 0121...