21
Operations & Maintenance Manual: Setting Up
PROCEDURE 4. ADJUST USER PREFERENCES
On the Preference screen, various BVI 3000 operating parameters can be customized for personalized use. The
following table documents the configurable parameters and their available values.
Table 3.
User Preferences
PARAMETER
RANGE OR VALUE
DESCRIPTION
Backlight
AUTO (default)
The backlight turns on and off automatically, depending
on the ambient light conditions
ON
The backlight is always on
OFF
The backlight is always off
Beep volume
0–9 (default is 5)
Adjusts the instrument volume level up (higher values) or
down (lower values)
Language
Multiple languages
available
Specifies the language used for displays and printouts
Date
MM/DD/YY
Month, day, year
DD/MM/YY
Day, month, year
YY/MM/DD
Year, month, day
Printout
ADD_WALLS (default)
Print grayscale, B‑mode images with bladder walls
highlighted
NO_IMAGES
Do not print any B‑mode images
RAW_ONLY
Print grayscale B‑mode images without identifying the
bladder walls
WALLS_ONLY
Bladder outline only
12_PLANES
All 12 scan planes, printed as in the ADD_WALLS mode
Quick
OFF (default)
When you press the PRINT button, the Print screen
appears, and you may select whether you would like to
print the last scan result, cost savings, or a test print.
ON
When you are on the Scan Results screen, if you press
the PRINT button, the scan results print automatically (the
Print screen does not appear).
UTI rate
0–100% (default is 3%)
Percentage of catheterizations that result in UTI; used in
cost‑savings calculations
UTI cost
0‑2000 (default is 680)
Cost to treat a UTI; used in cost‑savings calculations
Cath cost
0–20 (default is 3)
Cost of a catheter; used in cost‑savings calculations
Cath vol
0–1000 ml (default is 300
ml)
Bladder volume below which catheterization is
unnecessary; used in cost‑savings calculations
Flash
OFF (default)
Scan results are not saved if instrument is turned off;
however, processing time is slightly faster
ON
Flash memory is enabled, and scan results are saved
even the instrument turns off
Time zone
GMT ± 0–12 (default is 0) This function is currently not used